background image

3.1 GUI Installations

Installation instructions follow for GUI control of the SPECTRA. Connectivity between the computer and your 
SPECTRA can be accomplished one of two ways; either using a RS-232 straight-thru cable or using the optional 
USB-to-RS232 adapter cable. Both methods are covered below.

GUI Installation and Operating Instructions (using RS-232 straight-thru cable)

Download the zip file for the SPECTRA GUI from 

http://www.lumencor.com/software_control.html

Unzip the file and run setup.exe to install the SPECTRA GUI.

Connect the RS-232 cable between the PC and the SPECTRA.

Connect the power supply to the SPECTRA.

Toggle the power switch on the rear panel to the ON “1” position. The green LED next to the switch should light.

Run the GUI by going to the Program Menu and selecting LLE Controller.

In the GUI, select “COMPUTER” mode.

In the COM pulldown menu, select the COM port assigned to serial communications. Generally this is COM1. [If 
you are unsure which one that is then go to Control Panel, then System, then the Hardware tab. Select Device 
Manager to see the hardware profile. Expand “PORTS (COM & LPT)” to see which COM port is assigned to the 
“Communications Port” and select that port in the GUI.] The PC should now have control of each color channel in 
the SPECTRA. You will have the ability to turn each color ON or OFF and adjust the intensity of each individually. 
THIS STEP IS REQUIRED TO RE-ESTABLISH THE COMMUNICATION LINK WHENEVER THE SPECTRA LIGHT 
ENGINE IS POWER CYCLED OR WHENEVER THE GUI IS STARTED.

GUI Installation and Operating Instructions (using USB-to-RS232 cable)

Download the zip file for the SPECTRA GUI from 

http://www.lumencor.com/software_control.html

Unzip the file and run setup.exe to install the SPECTRA GUI.

Connect the USB-to-RS-232 cable between the PC and the SPECTRA.

Connect the power supply to the SPECTRA.

Toggle the power switch on the rear panel to the ON “1” position. The green LED next to the switch should light.

Run the GUI by going to the Program Menu and selecting LLE Controller.

In the GUI, select “COMPUTER” mode.

In the COM pulldown menu, select the COM port assigned to USB-to-Serial communications. [If you are unsure 
which one is correct then go to Control Panel, then System, then the Hardware tab. Select Device Manager to see 
the hardware profile. Expand “PORTS (COM & LPT)” to see which COM port is assigned to the “USB-to-Serial 
Comm Port” and select that port in the GUI.] The PC should now have control of each color channel in the 

SPECTRA, SPECTRA S and SPECTRA X Instruction Manual

8

Summary of Contents for SPECTRA Light Engine

Page 1: ...Lumencor SPECTRA Light Engine Instruction Manual Lumencor Inc 14964 NW Greenbrier Parkway Beaverton OR 97006 T 503 213 4269 F 503 536 6741 www lumencor com...

Page 2: ...ent is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmf...

Page 3: ...gs 3 Installation and Operating Instructions 4 Spectral Output 5 Routine Maintenance and Trouble Shooting 6 Customer Support 7 Product Specifications 8 Connectors 9 Declaration of Conformance 10 Warra...

Page 4: ...on the rear panel to turn on off the power into the unit A green power indicator on the rear cover is lit ON when the power supply is connected to the SPECTRA and the power switch is in the on positio...

Page 5: ...ion du SPECTRA ne comporte pas de fusible Le mat riel doit tre aliment par une source d alimentation continue qui a t d clar e conforme aux crit res d un circuit d nergie limit e en vertu de la clause...

Page 6: ...al de toutes les mesures de s curit et mises en garde affich es 3 Installation and Operating Instructions The SPECTRA family of light engines all ship with the following list of standard components 1...

Page 7: ...filters There are fixed in the SPECTRA and SPECTRA S The filters are installed in individual holders which the user can access once the top cover retained by two thumb screws is removed Refer to the...

Page 8: ...r be controlled by a Lumencor supplied GUI via a RS 232 port by the Lumencor Remote Control Accessory RCA via the RS 232 port or by a 3rd party laboratory software program via the TTL port The GUI pro...

Page 9: ...ons Port and select that port in the GUI The PC should now have control of each color channel in the SPECTRA You will have the ability to turn each color ON or OFF and adjust the intensity of each ind...

Page 10: ...cable between the PC and the TTL port on the rear of the SPECTRA The integration and set up of the 3rd party laboratory software is very application specific and beyond the scope of this document For...

Page 11: ...or accumulation on the air intake ports A vacuum may be used to remove debris so that a steady supply of air is available for cooling It is recommended that these dust filters be cleaned by a gentle s...

Page 12: ...erating 0 to 10 000 feet 3 048 meters Non operating 0 to 20 000 feet 6 096 meters Dimensions LxWxH 11 0 x 7 0 x 4 0 in 27 9 x 17 8 x 10 2 cm Weight 9 lb 4 1 kg Lifetime 10 000 hr Input Power Requireme...

Page 13: ...frequency counter It outputs a square wave 0 5V frequency is proportional to the power of the light being measured SPECTRA SPECTRA S and SPECTRA X Instruction Manual 12 Pins Definition DC Characterist...

Page 14: ...that the SPECTRA Light Engine conforms to the following directives and norms EMC directive 2004 108 EC 2006 95 EC CB Scheme CE Declaration of Conformity TUV NRTL Listing TUV Canadian Listing TUV Europ...

Reviews: