background image

1.  Introduction

The Lumencor SPECTRA family of light engines (SPECTRA, SPECTRA S and SPECTRAX) are designed for laboratory 
use by bioanalytical researchers and/or developers of life science instrumentation. The SPECTRA (all models) provides 3, 
4, 5, 6 or 7 spectrally discrete, bright, controllable light outputs directly to a sample; or in the case of fluorescence 
microscopy, to the objective. Each of the colors is produced by an independent module that has been optimized to 
produce a precise set of wavelengths.

This manual covers all SPECTRA models: SPECTRA X-YYY-ZZ or SPECTRAX X-YYY-ZZ. X denotes the number of 
colors, YYY denotes a unique customer code and ZZ denotes option and revision. The first Z variable indicates a 
standard SPECTRA if it is blank or a color mix or simultaneous output SPECTRA (or SPECTRA X) if it is an “S”. The 
second Z denotes the revision level. Any reference to “SPECTRA” in this manual shall apply to all SPECTRA models 
unless noted otherwise.

The light sources within the SPECTRA are controlled by software; either via a serial interface (RS-232 or USB) to a 
computer running a Lumencor supplied GUI or via a TTL interface to a computer running a third party microscopy 
software application.The user can enable or disable each source independently and change the intensity of each source 
independently. The only manual control is a power switch on the rear panel to turn on/off the power into the unit. A green 
power indicator on the rear cover is lit (“ON”) when the power supply is connected to the SPECTRA and the power 
switch is in the on position.

2.  Precautions and Warnings {Précautions et mises en garde}

A few simple practices will ensure trouble-free operation for the life of the light engine.

Les quelques règles simples suivantes permettront d’assurer un fonctionnement fiable pendant toute la durée de service 
de la source lumineuse.

Safety Instructions:

Please read and follow all safety instructions provided 

BEFORE

 using your new SPECTRA. Failure to comply with the 

safety instructions may result in fire, electrical shock, or personal injury and may damage or impair protection provided by 
equipment. Please save all safety instructions.

Instructions de sécurité

Veiller à lire et à respecter toutes les instructions de sécurité fournies 

AVANT

 d’utiliser le nouveau SPECTRA afin d’écar-

ter les risques d’incendie, de décharge électrique, de blessure corporelle et de possibles dommages ou défaillance de la 
protection offerte par l’appareil. Conserver toutes les instructions de sécurité.

Safety Definitions 

{

Définitions relatives à la sécurité}

Warning: 

Statements identify conditions or practices that could result in personal injury.

Avertissement

:

 déclarations qui identifient des situations ou des pratiques susceptibles d’entraîner des blessures cor-

porelles

Caution: 

Statements identify conditions or practices that could result in damage to your equipment. 

SPECTRA, SPECTRA S and SPECTRA X Instruction Manual

3

Summary of Contents for SPECTRA Light Engine

Page 1: ...Lumencor SPECTRA Light Engine Instruction Manual Lumencor Inc 14964 NW Greenbrier Parkway Beaverton OR 97006 T 503 213 4269 F 503 536 6741 www lumencor com...

Page 2: ...ent is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmf...

Page 3: ...gs 3 Installation and Operating Instructions 4 Spectral Output 5 Routine Maintenance and Trouble Shooting 6 Customer Support 7 Product Specifications 8 Connectors 9 Declaration of Conformance 10 Warra...

Page 4: ...on the rear panel to turn on off the power into the unit A green power indicator on the rear cover is lit ON when the power supply is connected to the SPECTRA and the power switch is in the on positio...

Page 5: ...ion du SPECTRA ne comporte pas de fusible Le mat riel doit tre aliment par une source d alimentation continue qui a t d clar e conforme aux crit res d un circuit d nergie limit e en vertu de la clause...

Page 6: ...al de toutes les mesures de s curit et mises en garde affich es 3 Installation and Operating Instructions The SPECTRA family of light engines all ship with the following list of standard components 1...

Page 7: ...filters There are fixed in the SPECTRA and SPECTRA S The filters are installed in individual holders which the user can access once the top cover retained by two thumb screws is removed Refer to the...

Page 8: ...r be controlled by a Lumencor supplied GUI via a RS 232 port by the Lumencor Remote Control Accessory RCA via the RS 232 port or by a 3rd party laboratory software program via the TTL port The GUI pro...

Page 9: ...ons Port and select that port in the GUI The PC should now have control of each color channel in the SPECTRA You will have the ability to turn each color ON or OFF and adjust the intensity of each ind...

Page 10: ...cable between the PC and the TTL port on the rear of the SPECTRA The integration and set up of the 3rd party laboratory software is very application specific and beyond the scope of this document For...

Page 11: ...or accumulation on the air intake ports A vacuum may be used to remove debris so that a steady supply of air is available for cooling It is recommended that these dust filters be cleaned by a gentle s...

Page 12: ...erating 0 to 10 000 feet 3 048 meters Non operating 0 to 20 000 feet 6 096 meters Dimensions LxWxH 11 0 x 7 0 x 4 0 in 27 9 x 17 8 x 10 2 cm Weight 9 lb 4 1 kg Lifetime 10 000 hr Input Power Requireme...

Page 13: ...frequency counter It outputs a square wave 0 5V frequency is proportional to the power of the light being measured SPECTRA SPECTRA S and SPECTRA X Instruction Manual 12 Pins Definition DC Characterist...

Page 14: ...that the SPECTRA Light Engine conforms to the following directives and norms EMC directive 2004 108 EC 2006 95 EC CB Scheme CE Declaration of Conformity TUV NRTL Listing TUV Canadian Listing TUV Europ...

Reviews: