6.  Routine Maintenance and Trouble Shooting

 

No routine maintenance is required.  There are no user-replaceable components or sub-assemblies in the SOLA SM light 
engine.  Opening the light engine enclosure will void the manufacturer’s warranty.  In the event that the light engine fails to 
perform in accordance with the specifications listed on the certificate of conformance, please contact Lumencor 
Technical Support for assistance, as directed in Section 7. 

7.  Customer Support

For technical support of SOLA SM and SOLA SM 365 light engines, please contact Lumencor by phone at 
503.213.4269 or via e-mail at 

[email protected].

  Please be prepared to provide the 4- or 5-digit serial number 

of the light engine (see Figure 1), a description of the problems encountered and information on the usage context (e.g. 
what microscope is being used).  This information will help to determine whether the problems can be resolved in situ by 
adjustments to the system configuration, or whether a fault has developed in the light engine that requires its return to 
Lumencor’s facility in Beaverton, Oregon for evaluation and, if necessary, repair.  Any light engine return to Lumencor for 
service or repair requires a material authorization (RMA) number.  To request an RMA number, fill out and submit the 

online request form

. It is the customer’s responsibility to properly package and safely ship products to Lumencor.  

Instructions for shipping will be provided in the e-mail giving notification of the RMA number. 

8.  Warranty

SOLA SM and SOLA SM 365 light engines come with a 18 month warranty, starting on the original date of shipment from 
Lumencor.  Light Engines qualifying for warranty service must be verifiably delivering performance that is substantially at 
variance with the levels documented in the certificate of conformance.  The light engine must also have been used and 
maintained under operating conditions consistent with the specifications given in Section 5, and observing all the 
Precautions and Warnings notified in Section 2.  This warranty does not extend to light engines that have been subject to 
misuse, accident, tampering or improper installation.  Accessories including (but not limited to) liquid light guides, optical 
fibers, collimators, cables and control consoles are not covered by the warranties attached to light engines. Please fill out 
and submit the 

online warranty registration form 

.  This will facilitate provision of warranty service should it be required. 

Dimensions (W x L x H)

4.9 in x 10.4 in x 6.40 in  /  12.5 cm x 26.3 cm x 16.3 cm

Weight

8.0 lb  /  3.6 kg

Input Power Requirements

24 V DC / 5.0 A, 120 W maximum, power supply included

Power consumption

90 W (light output on); 3 W (light output off, fan on)

Warm-up Period

1 s

Protection

IP Rating of X0

Sound Level

Sound Level at 1 m < 10 db(A)

Connections

3.5 mm foot switch (optional accessory)

Warranty

18 months parts and labor

Specification

Detail

SOLA SM Light Engine Instruction Manual 

 

11

Summary of Contents for SOLA SM

Page 1: ...SOLA SM and SOLA SM 365 Light Engine Instruction Manual Lumencor Inc Document Number 57 10003 www lumencor com ...

Page 2: ...10 CE Marking Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU Declarations of Conformity can be found at https lumencor com company compliance For EU customers discarding end of life Lumencor electrical and electronic equipment Please submit an RMA request with Recycle product under WEEE in the Description of Is...

Page 3: ...ion 2 Precautions and Warnings 3 Installation and Operating Instructions 4 Spectral Output 5 Product Specifications 6 Routine Maintenance and Trouble Shooting 7 Customer Support 8 Warranty SOLA SM Light Engine Instruction Manual 3 ...

Page 4: ...lid state light sources The light output on off status is manually controlled via a rocker switch located on the front panel or a foot pedal toggle switch accessory part number 29 10045 that plugs into the 3 5 mm connector on the rear panel This manual covers both SOLA SM and SOLA SM 365 light engine models which are defined by their different ultraviolet violet spectral outputs as described in Se...

Page 5: ...de sécurité Warning DO NOT use an unapproved power supply The Lumencor supplied external power supply is recommended for use with the SOLA SM light engine Alternative approved certified 24 V DC power supplies may be used provided that the current is limited to 5 0 A max Also it is imperative that the alternative power supply has output over current protection as the power input of the SOLA SM is n...

Page 6: ...iq uides animaux ou végétaux RISK GROUP 3 Warning Possibly hazardous optical radiation emitted from this product Do not look at operating lamp Eye injury may result Warning Infrared IR emitted from this product Do not look at operating lamp Warning UV emitted from this product Avoid eye and skin exposure to unshielded product GROUPE DE RISQUE 3 Avertissement Infrarouge IR émise par ce produit Ne r...

Page 7: ... désalignés par le choc résultant d une chute sur une surface dure DISCLAIMER Lumencor shall not be liable for injury to the user or damage to the product resulting from the SOLA SM light engine being used in a way for which it was not intended and in complete disregard for any posted safety precautions and warnings AVIS DE NON RESPONSABILITÉ Lumencor décline toute responsabilité pour les blessure...

Page 8: ...it on a hard surface and avoid blocking or restricting airflow at the air intake front panel and exhaust ports rear panel Figure 1 The light engine must be operated in a non condensing environment dew point 10ºC with controlled ambient temperature 30º C Restricting the airflow will cause the unit to operate at elevated temperatures and will result in decreased operating lifespan and or premature f...

Page 9: ...rmation resulting in decreased light transmission Take necessary precautions to protect yourself and others from the high intensity light when turning on the unit 3 3 Operation The SOLA SM light output can be turned on or off using the manual light output switch on the front panel or using the foot pedal toggle switch accessory Do not combine the operation of these switches If the front panel rock...

Page 10: ...engine is sold or transferred the certificate of conformance should be conveyed to the new owner Certificates of conformance are also recorded in Lumencor s database and copies can be requested by e mail to techsupport lumencor com The request message must include the 4 or 5 digit serial number of the light engine Specification Detail Temperature Operating 32 to 95 F 0 to 35 C Non operating 4 to 1...

Page 11: ...y ship products to Lumencor Instructions for shipping will be provided in the e mail giving notification of the RMA number 8 Warranty SOLA SM and SOLA SM 365 light engines come with a 18 month warranty starting on the original date of shipment from Lumencor Light Engines qualifying for warranty service must be verifiably delivering performance that is substantially at variance with the levels docu...

Reviews: