conformance for reference.  In the event that the light engine is sold, the certificate of conformance 
should be transferred to the new owner. 

 

3.2  Installation

When setting up the LIDA light engine, be sure that the adapter is 
securely attached to the appropriate illumination port of the 
microscope.  Usually this will be the transmitted light port (Figure 2).  
Be careful to properly support the LIDA light engine until it is 
mechanically coupled to the instrument. 

The LIDA light engine should be operated in an open area where 
airflow facilitates proper cooling.  Restricting airflow around the unit 
will result in elevated operating temperatures and could result in 
decreased service life and/or premature failure. 

With the manual control knob (Figure 3) in the OFF position, 
plug the DC power supply into the power connector at the 
back of the light engine.  A green indicator light on the rear 
panel above the DC power input connector indicates that 
electrical power is being supplied to the light engine. 

3.3  Operation: Manual Control

Take necessary precautions to protect yourself and 
others from the high intensity light when turning on 
the unit.  The LIDA light engine should be coupled to a 
microscope before it is turned on

.  

Two control knobs 

are located on the side panel (Figure 3).  The CONTROL 
knob turns white light output on or off [1].  The 
yellow “light” indicator on the side panel alerts the 
user when the light output is on.  In the “ext” 
position, control is transferred to the USB (“external”) 
port on the rear panel (Figure 4) for control pod 
(Section 3.4) or computer workstation (Section 3.5) 
operation.  The INTENSITY control knob has four positions 
that set the light engine to deliver approximately 10%, 25%, 
50% and 100% of maximum output.  There is no warm-up time; 
the light engine output stabilizes less than 1 second after the 
CONTROL knob is turned to the “on” position.  Light output can be switched off during intervals when it is 
not required for active viewing or data collection.

Notes 

[1] Manual controls do not allow selection of the individual red, green and blue component light sources.  
This functionality is accessed through the serial (Sections 3.4  and 3.5) or TTL (Section 3.6) interfaces. 

LIDA Light Engine Instruction Manual

7

Figure 2.

  

LIDA light engine with output 

adapter for coupling to Nikon Ti/Ni/Ci 
microscope transmitted light port.

Set screw 
fastener

Locating pin

Figure 1.

 Model and serial number 

identification label on the LIDA light 
engine rear panel.

Summary of Contents for LIDA Light Engine

Page 1: ...LIDA Light Engine Instruction Manual Lumencor Inc Document Number 57 10005 www lumencor com...

Page 2: ...an radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Safety Certifications TUV SUD America C...

Page 3: ...oduction 2 Precautions and Warnings 3 Installation and Operating Instructions 4 Spectral Output 5 Specifications 6 Routine Maintenance and Troubleshooting 7 Customer Support 8 Warranty LIDA Light Engi...

Page 4: ...rvice de la source lumineuse Safety Instructions Please read and follow all safety instructions provided BEFORE using your new LIDA light engine Failure to comply with the safety instructions may resu...

Page 5: ...ioanalytjcal instrument Avertissement NE PAS regarde directement la sortie de la source lumineuse L intensit lumineuse de cette source est sup rieure celle de la majorit des appareils d clairage dispo...

Page 6: ...product resulting from the LIDA light engine being used in a way for which it was not intended and in complete disregard for all posted safety precautions and warnings AVIS DE NON RESPONSABILIT Lumen...

Page 7: ...ng on the unit The LIDA light engine should be coupled to a microscope before it is turned on Two control knobs are located on the side panel Figure 3 The CONTROL knob turns white light output on or o...

Page 8: ...ay screen 3 Press the left button to select the desired color channel Successive presses will cycle through the available color channels 4 Press the right button to turn the selected light source on A...

Page 9: ...ces Manage Devices Add and select Lumencor LIDA The serial port connection dialog box will appear Select the COM port connected to the LIDA light engine 2 6 Click Configure Device to open the Lida Con...

Page 10: ...ng temperature of the LIDA on the LIDA Control Pad Figure 7 4 Camera model according to selection made in step 4 3 6 TTL Control TTL provides light output on off switching but not intensity control Ho...

Page 11: ...p Therefore on off controls in the GUI or other serial control software must be set in the OFF state when using TTL triggering 4 Spectral Output The typical spectral output distribution of a LIDA ligh...

Page 12: ...tive humidity non condensing Altitude Operating 0 to 10 000 feet 3 048 meters Non operating 0 to 45 000 feet 13 176 meters Dimensions Size WxLxH 4 25 in x 6 0 in x 4 25 in 10 5 cm x 15 cm x 10 5 cm We...

Page 13: ...requires its return to Lumencor s facility in Beaverton Oregon for evaluation and if necessary repair Any light engine return to Lumencor for service or repair requires a material authorization RMA nu...

Reviews: