background image

using the master power switch.  The embedded green indicator will illuminate (if applicable, ensure 
that the liquid light guide is inserted and secured; see below).

The AURA Light Engine may be configured for light delivery via a liquid light guide (Figure 3) or an 
SMA-terminated optical fiber.  The output couplers are built into the light engine front panel and are 
not field-exchangeable.  The liquid light guide output coupler incorporates an electrical power 
interlock switch in the back of the receptacle.  The liquid light guide must be fully inserted in the 
receptacle to activate the interlock.  Electrical power supply to the light engine cannot be turned on 
via front panel power switch until the interlock is activated.  After fully inserting the liquid light guide, 
lock it in position using the set screw on the side of the receptacle to prevent inadvertent 
disconnection of the light guide during use.

3.3  Operation 

3.3.1 Control and Interlock Keys 

The 

Master Power Switch

 button on the front panel (Figure 3) turns the electrical power to the unit 

on or o

.  A green power indicator embedded in the button is lit when the power supply is 

connected to the light engine and the power button is in the on position.  Initialization of the onboard 
computer takes about 30 seconds after the master power switch is turned on.  When initialization is 
complete, the status indicator display (Figure 3) will activate.

The 

Key Control

 (Figure 3) must be in the on position before light output can be turned on.  The key 

must  be removed and stored in a secure location when the product is not in use.  

ONLY

 trained 

individuals should use and have access to the key.  The 

Master Power Switch

 button, 

Key control

 

and 

Remote interlock

 (Figure 2) can be used to shut o

 light output.

The 

Source Enabled

 indicator LED (below the status indicator display; Figure 3) provides a warning 

indication that one or more light sources are active and emitting invisible and/or visible radiation.

The 

Remote Interlock Connector

: (Figure 4) is provided to allow for connection of a remote 

interlock.  When this interlock is open it will shut o

 light output.  After the interlock has been 

opened, the 

Manual Reset

 button will need to be pushed to resume light output.

WARNING:

 

The Remote Interlock relies on a passive continuity circuit for proper operation.

 

 It does 

not provide a voltage source for external circuitry, nor can external voltage sources be applied to it. 

UNDER NO

 

CIRCUMSTANCES

 should a voltage or voltage source be applied to the Remote 

Interlock circuit.

  

APPLYING A VOLTAGE OF EITHER POLARITY MAY RESULT IN ANOMALOUS 

OPERATION AND/OR DAMAGE THE PRODUCT, AND COULD IMPACT SAFETY. 

AURA Light Engine Manual 57-10011

10

Summary of Contents for AURA Light Engine

Page 1: ...Lumencor Light Engine Manual AURA Light Engine AURA Light Engine Manual 57 10011 lumencor com...

Page 2: ...ication EN 61010 1 2010 CE Marking Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU Declarations of Conformity can...

Page 3: ...Precautions and Warnings 3 Installation and Operating Instructions 4 Light Output Characteristics 5 Operational Specifications 6 Routine Maintenance and Trouble Shooting 7 Customer Support 8 Warranty...

Page 4: ...utput over time The light sources within the AURA Light Engine are controlled by an on board microprocessor operating Lumencor firmware accessed via one of two serial interfaces USB RS 232 or TCP The...

Page 5: ...ce de la protection o erte par l appareil Conserver toutes les instructions de s curit Safety Definitions D finitions relatives la s curit Warning Statements identify conditions or practices that coul...

Page 6: ...t sup rieure celle de la majorit des appareils d clairage disponibles dans le commerce et est con ue pour un raccordement direct un microscope ou autre appareil de bioanalyse Warning DO NOT turn on th...

Page 7: ...may damage your light engine Attention NE PAS placer de liquide sur la source lumineuse Les liquides renvers s peuvent endommager la source lumineuse Caution DO NOT drop the light engine It contains g...

Page 8: ...lock jumper Figure 2 6 Quickstart Guide instruction document 85 10049 AURA Light Engine Manual 57 10011 8 AC Power Cords Region Part Number North America 29 10002 Europe 29 10005 United Kingdom 29 100...

Page 9: ...he AURA Light Engine must be operated in a non condensing environment dew point 10 C with controlled ambient temperature 30 C Thermal overload protection is provided by the on board computer in conjun...

Page 10: ...oard computer takes about 30 seconds after the master power switch is turned on When initialization is complete the status indicator display Figure 3 will activate The Key Control Figure 3 must be in...

Page 11: ...GE DISTANCE PEUT ENTRA NER UN COMPORTEMENT ANOMAL ET OU ENDOMMAGER LES PRODUITS WARNING Prior to turning the light output on be sure the output end of the liquid light guide or optical fiber is safely...

Page 12: ...Ensure that the distal end of the light guide or optical fiber is safely directed into an enclosed optical path e g microscope epi illuminator or a beam dump before turning the light output on If the...

Page 13: ...tal factors in photometric and quantitative imaging applications where reproducibility and accuracy are essential To use power regulation a desired power reference value in milliwatts is entered in th...

Page 14: ...AURA Light Engine control mode change the setting by holding down the MODE button on the pod until the light engine selection menu appears Move the cursor to AURA in the selection menu by turning the...

Page 15: ...y on or by turning on additional sources commands will be rejected To clear the error condition reduce intensities of sources that are on or turn o additional sources Light source outputs are refined...

Page 16: ...ate of conformance for reference In the event that the light engine is sold the certificate of conformance should be transferred to the new owner Certificates of conformance are also recorded in Lumen...

Page 17: ...to 10 000 feet 3 048 meters Non operating 0 to 45 000 feet 13 176 meters Dew Point Operating 32 to 59 F 0 to 15 C Dimensions Size W x L x H 145 mm x 263 mm x 203 mm 5 7 in x 10 4 in x 8 0 in Weight 6...

Page 18: ...uide must be inserted in light output receptacle 3 Remote interlock jumper must be inserted in rear panel 4 External gate jumper must be inserted in front panel Figure 3 Unusually weak fluorescence si...

Page 19: ...onditions consistent with the specifications given in Section 5 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subje...

Reviews: