INFRATHERM

 pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO 

 

 

38

(fixed socket:  

model ERA.0S.302.CLL, 

straight plug: 

model FFE.0S.302.CLAC50    

Fa. Lemosa GmbH, http://www.lemo.de)

 

 

Pin 1 

Video-Output: BAS-Signal (white) 

 

Cinch: middle pin  SCART: Pin 20 

 

 

Pin 2 

Video- Output: Ground (brown) 

 

Cinch: shield 

SCART: Pin 17 

 

 

 

Note:  

The video output is not isolated from the power supply, therefore if several units are 
switched together using a video multiplexer, then each unit must have its own iso-
lated power supply. 

 
 

4.2 

Connecting the pyrometer to a PC 

The pyrometer is equipped with a serial interface RS232 or RS485. Standard on a PC is the RS232 inter-
face. At this interface one pyrometer can be connected if the interface is set to RS232. Only short distances 
can be transmitted with RS232 and electromagnetic interferences can affect the transmission.  
With RS485 the transmission is to a large extend free of problems, long transmission distances can be real-
ized and several pyrometers can be connected in a bus system. If RS485 is not available at the PC, it can be 
realized with an external converter which converts the RS485 in RS232 for a standard connection to a PC.  
When using a converter RS485 

Ù

 RS232 take care, that the converter is fast enough to receive the pyrome-

ter’s answer to an instruction of the master. Most of the commonly used converters are too slow for fast 
measuring equipment. So it is recommended to use the IMPAC-converter (order no. 3 852 430). 
 
 

4.3 

Connection via interface RS232 

The transmission rate (in baud) of the serial interface is 
dependent on the length of the cable

Values between 

1200 and 38400 Bd may be set.  
The baud rate has to be reduced by 50% when the 
transmission distance is doubled (see also 

9.4 Bau

rate

).  

Typical cable length for RS232 at 19200 Bd is 7 m. 
The software 

InfraWin 

contains a function to test the 

reliability of the transmission with the selected baud 
rate (see under 

10.7

,

 Basic settings 

Æ

 Test

)). 

PC

K

A

B

C

D

J

G

F

E

H

M

L

1

 

  2 3 4 5

6 7 8 9

 DGND (red)

 

 RxD (black)

 

 TxD (violet)

 

Pyrometer

 
 

4.4 

Connection via interface RS485 

Half-duplex mode: 
A1 and A2 as well as B1 and B2 are 
bridged in the 12-pin round connector of 
the connecting cable, to prevent reflec-
tions due to long stubs. It also safeguards 
against the interruption of the RS485 Bus 
system should a connecting plug be 
pulled out. The master labels mark the 
connections on the RS485 converter. The 
transmission rate of the serial interface in 
Baud (Bd) is dependent on the length of 
the cable

Values between 1200 and 

38400 Bd may be set. 
The standard cable length for 19200 Bd is 
2 km. The baud rate is reduced by 50% 
when the transmission distance is doubled 
(see 

9.4 Baud rate

). 

 

Terminator 120 Ohm

A
B
S

K

A

B

C

D

J

G

F

E

H

M

L

A

B

C

D

F

E

L

K

A

J

G

F

H

M

L

D

G

ND 

B1

 

A2

 

B2

 

A1

 

D

G

ND 

B1

 

A2

 

B2

 

A1

 

DGND

 

A2

 

B2

 

A1

 

B1

 

Master

Pyrometer 32 

e.g. address 31

Pyrometer 2 

e.g. address 01

 

Pyrometer 1 

e.g. address 00

 

 
 
 

Summary of Contents for Impac IGA 5

Page 1: ...IS 5 IGA 5 IS 5 LO IGA 5 LO INFRATHERM Pyrometer IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry Betriebsanleitung User Manual...

Page 2: ...S485 38 4 5 Connection schematic for analysing devices 39 5 Mechanical installation 39 5 1 Accessories option 39 5 2 Connection overview IS 5 LO IGA 5 LO 40 6 Instrument alignment 41 6 1 View finder 4...

Page 3: ...48 10 8 Configuration of the display on the TV screen 49 10 8 1 The PID controller 50 10 8 2 Self tuning algorithm 51 10 9 Measurement color bar 51 10 10 Measurement online trend 52 10 11 Output listi...

Page 4: ...otify any person or persons of such revisions or changes All series 5 Instruments from IMPAC Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects...

Page 5: ...isolated from each other Parameters Adjustable on the pyrometer Emissivity exposure time t90 analog output 0 20 or 4 20 mA online offline switch for adjustments at pyrometer or PC Additionally adjusta...

Page 6: ...levels controlled by the measuring temperature Field of view approx 10 x 14 of focused distance Date time real time clock with at least 3 days spare run Video output plug separate round plug at the p...

Page 7: ...if the person already used similar instruments or was already trained by the manu facturer The pyrometer has only to be used for the purpose described in the manual It is recommended to use only acces...

Page 8: ...e via external switch J pink see 4 1 1 external clearing of max value storage or hold function Only pyrometers with PID controller controller activated set controller output to 0 emergency stop G red...

Page 9: ...r fast measuring equipment So it is recommended to use the IMPAC converter order no 3 852 430 4 3 Connection via interface RS232 The transmission rate in baud of the serial interface is dependent on t...

Page 10: ...made from stainless steel protects the pyrometer if exposed to a hot environment It is designed for ambient temperatures up to 180 C Water cooling jacket with integrated air purge Miscellaneous for IS...

Page 11: ...way to align the pyrometer to the measured object The clamping screws of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions Mounting angle Ball and sock...

Page 12: ...Attention Do not look directly into the laser beam Laser class 2 according to IEC 60825 1 3 4 Note The laser warning signs on the pyrometer should be easily viewable at all times even after it has bee...

Page 13: ...2 15 7 2 Differing to the focused measuring distance all instrument types The optical profiles show that the spot size M varies with the distance from the pyrometer to the measuring object Spot sizes...

Page 14: ...out bending or twisting it Caution Disconnect the cable before opening the cover When reassembling the cover insert it carefully into the guide pin and then fasten it with the screws After that the co...

Page 15: ...r a correct measurement it is necessary to adjust the emissivity This emissivity is the relationship between the emission of an real object and the emission of a black body radiation source this is an...

Page 16: ...ttings The digital PC interface enables the exchange of data with a PC either by using the InfraWin software sup plied or an user adapted communication software see command table in section 14 Data fo...

Page 17: ...stab lished after each impulse from the reset signal double storage when entering the reset intervals via push buttons or PC interface the double storage is automatically selected This mode utilizes t...

Page 18: ...n of the individual icons in the program s help menu Click on the F1 button after InfraWin has been loaded or click on the in the menu bar The following descriptions refer to program version 4 0 The l...

Page 19: ...of devices InfraWin changes to the display of 1 or 2 devices If 2 devices are selected always 2 windows are displayed for settings or evaluation 10 7 Basic settings Under pyrometer parameters all pre...

Page 20: ...indow the connection with the preset baud rate can be checked Here the command was send 500 times with 19200 baud It has taken 4 56 seconds and no transmission errors has occurred 10 8 Configuration o...

Page 21: ...sired temperature a corresponding signal which changes over the time is transmitted to the analog output until the deviation reaches zero The integral time can be set in two different formats min s fo...

Page 22: ...red temperature Problems are generally posed by directly controlled systems where there are high contrasts between the cooling time constant and the heating up time constant For example in many cases...

Page 23: ...cally saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old values are over written when a new measurement is taken Files from older pr...

Page 24: ...er file the previous file will be overwritten and replaced by the standard i12 file 10 14 PC sampling rate time interval between two measurements This function sets a time interval After each interval...

Page 25: ...lower correct emissivity corresponding to the material see 8 5 Lens contaminated Clean lens carefully see 12 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corre...

Page 26: ...repeated AAmsXXX XXX 000 999 auto repeat of the ms command Reading basic temperature range AAmb Output XXXXYYYY 2 x 4 hexdigit for MBA and MBE C or F Reading sub range AAme Output XXXXYYYY 2 x 4 hexdi...

Page 27: ...rtional control Reset time Integral time AATiXX XX AATiXX XX or XX XX Ti time in format min sec XX XX is equivalent to 00 01 99 99 calculation is every second Ti time in Format XX XX sec is equivalent...

Page 28: ...r RS485 IS 5 MB 20 600 2000 C 3 857 150 3 857 170 3 857 160 3 857 180 IS 5 MB 25 800 2500 C 3 857 200 3 857 220 3 857 210 3 857 230 IS 5 MB 30 1000 3000 C 3 857 250 3 857 270 3 857 260 3 857 280 IGA 5...

Page 29: ...justable for optical head I 3 820 330 Connection cable 5 m long straight pyrometer connector 3 820 500 Connection cable 10 m long straight pyrometer connector 3 820 510 Connection cable 15 m long stra...

Page 30: ...ent 41 Instrument settings 43 Interface commands 49 Interface connection 38 Internal temperature of the pyrometer 46 L Laser targeting light 36 41 M Maintenance 54 Maximum value storage 37 45 Measurem...

Page 31: ...IMPAC Infrared GmbH Temperaturmessger te Kleyerstra e 90 D 60326 Frankfurt Main Tel 49 0 69 9 73 73 0 Fax 49 0 69 9 73 73 167 Internet www impacinfrared com E Mail info impacinfrared com 3 869 207 a...

Reviews: