9.4 Baud 

rate 

The transmission rate of the serial interface in Baud (Bd) is dependent on the length of 
the cable. A standard cable length with RS232 for 19200 Bd is 7 m, with RS485 2 km. 
The baud rate is reduced by 50% if the transmission distance is doubled. 
 
 

9.5 

Wait time (tw) 

Using a pyrometer with RS485 it is possible that the connection is not fast enough to 
receive the pyrometer’s answer to a command of the master. In this case a minimum 
delay time (tw) can be set. The pyrometer waits this time until it answers a master in-
quiry (e.g.: tw = 02 at a baud rate 9600 means a wait time of 

2

/

9600

 sec). 

 
 

10 

Settings via interface and software 

 
The operating and analyzing software 

InfraWin 

is included in delivery of the pyrometer. With this software all 

pyrometer functions also can be used on the PC (except changing the interface or using the test current 
function). 
This section gives an overview about the functions of the software. Additionally there is a description of the 
individual icons in the program’s help menu. Click on the 

F1

 button after 

InfraWin

 has been loaded or click on 

the 

in the menu bar. 

The following descriptions refer to program version 4.0. The latest version is available for free as download 
from the homepage 

www.impacinfrared.com

 

 

 

Note:  

Before adjusting the parameters “emissivity”, “exposure time” and “analog output” via 
the software, the online 

Ù

 offline switch (DIP-switch Nr. 4) must be set to the “online” 

position in the pyrometer! 

 

 

10.1 

Connecting the pyrometer to a PC 

The program 

InfraWin 

can operate up to two devices. For two devices using the RS232 interface, two PC 

interfaces must be used. Two devices using RS485 may be operated simultaneously by the same interface, 
if two different addresses have been properly entered (see 

9.3 Address

). 

 
 

10.2 Installation 

For installation select the setup program „setup.exe“ from the 

InfraWin

-CD and follow the installation instruc-

tions. 
 
 

10.3 Program 

start 

After installation and the first program start a language must be chosen (German, English, French, Italian, 
Spanish. The language also can be changed in the program). On the start page the screen shows the follow-
ing icons: 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Settings: 

1.2 kBd 

. . . 

38.4 kBd 

Settings: 

00 Bit 

. . . 

99 Bit 

Summary of Contents for Impac IGA 5

Page 1: ...IS 5 IGA 5 IS 5 LO IGA 5 LO INFRATHERM Pyrometer IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry Betriebsanleitung User Manual...

Page 2: ...S485 38 4 5 Connection schematic for analysing devices 39 5 Mechanical installation 39 5 1 Accessories option 39 5 2 Connection overview IS 5 LO IGA 5 LO 40 6 Instrument alignment 41 6 1 View finder 4...

Page 3: ...48 10 8 Configuration of the display on the TV screen 49 10 8 1 The PID controller 50 10 8 2 Self tuning algorithm 51 10 9 Measurement color bar 51 10 10 Measurement online trend 52 10 11 Output listi...

Page 4: ...otify any person or persons of such revisions or changes All series 5 Instruments from IMPAC Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects...

Page 5: ...isolated from each other Parameters Adjustable on the pyrometer Emissivity exposure time t90 analog output 0 20 or 4 20 mA online offline switch for adjustments at pyrometer or PC Additionally adjusta...

Page 6: ...levels controlled by the measuring temperature Field of view approx 10 x 14 of focused distance Date time real time clock with at least 3 days spare run Video output plug separate round plug at the p...

Page 7: ...if the person already used similar instruments or was already trained by the manu facturer The pyrometer has only to be used for the purpose described in the manual It is recommended to use only acces...

Page 8: ...e via external switch J pink see 4 1 1 external clearing of max value storage or hold function Only pyrometers with PID controller controller activated set controller output to 0 emergency stop G red...

Page 9: ...r fast measuring equipment So it is recommended to use the IMPAC converter order no 3 852 430 4 3 Connection via interface RS232 The transmission rate in baud of the serial interface is dependent on t...

Page 10: ...made from stainless steel protects the pyrometer if exposed to a hot environment It is designed for ambient temperatures up to 180 C Water cooling jacket with integrated air purge Miscellaneous for IS...

Page 11: ...way to align the pyrometer to the measured object The clamping screws of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions Mounting angle Ball and sock...

Page 12: ...Attention Do not look directly into the laser beam Laser class 2 according to IEC 60825 1 3 4 Note The laser warning signs on the pyrometer should be easily viewable at all times even after it has bee...

Page 13: ...2 15 7 2 Differing to the focused measuring distance all instrument types The optical profiles show that the spot size M varies with the distance from the pyrometer to the measuring object Spot sizes...

Page 14: ...out bending or twisting it Caution Disconnect the cable before opening the cover When reassembling the cover insert it carefully into the guide pin and then fasten it with the screws After that the co...

Page 15: ...r a correct measurement it is necessary to adjust the emissivity This emissivity is the relationship between the emission of an real object and the emission of a black body radiation source this is an...

Page 16: ...ttings The digital PC interface enables the exchange of data with a PC either by using the InfraWin software sup plied or an user adapted communication software see command table in section 14 Data fo...

Page 17: ...stab lished after each impulse from the reset signal double storage when entering the reset intervals via push buttons or PC interface the double storage is automatically selected This mode utilizes t...

Page 18: ...n of the individual icons in the program s help menu Click on the F1 button after InfraWin has been loaded or click on the in the menu bar The following descriptions refer to program version 4 0 The l...

Page 19: ...of devices InfraWin changes to the display of 1 or 2 devices If 2 devices are selected always 2 windows are displayed for settings or evaluation 10 7 Basic settings Under pyrometer parameters all pre...

Page 20: ...indow the connection with the preset baud rate can be checked Here the command was send 500 times with 19200 baud It has taken 4 56 seconds and no transmission errors has occurred 10 8 Configuration o...

Page 21: ...sired temperature a corresponding signal which changes over the time is transmitted to the analog output until the deviation reaches zero The integral time can be set in two different formats min s fo...

Page 22: ...red temperature Problems are generally posed by directly controlled systems where there are high contrasts between the cooling time constant and the heating up time constant For example in many cases...

Page 23: ...cally saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old values are over written when a new measurement is taken Files from older pr...

Page 24: ...er file the previous file will be overwritten and replaced by the standard i12 file 10 14 PC sampling rate time interval between two measurements This function sets a time interval After each interval...

Page 25: ...lower correct emissivity corresponding to the material see 8 5 Lens contaminated Clean lens carefully see 12 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corre...

Page 26: ...repeated AAmsXXX XXX 000 999 auto repeat of the ms command Reading basic temperature range AAmb Output XXXXYYYY 2 x 4 hexdigit for MBA and MBE C or F Reading sub range AAme Output XXXXYYYY 2 x 4 hexdi...

Page 27: ...rtional control Reset time Integral time AATiXX XX AATiXX XX or XX XX Ti time in format min sec XX XX is equivalent to 00 01 99 99 calculation is every second Ti time in Format XX XX sec is equivalent...

Page 28: ...r RS485 IS 5 MB 20 600 2000 C 3 857 150 3 857 170 3 857 160 3 857 180 IS 5 MB 25 800 2500 C 3 857 200 3 857 220 3 857 210 3 857 230 IS 5 MB 30 1000 3000 C 3 857 250 3 857 270 3 857 260 3 857 280 IGA 5...

Page 29: ...justable for optical head I 3 820 330 Connection cable 5 m long straight pyrometer connector 3 820 500 Connection cable 10 m long straight pyrometer connector 3 820 510 Connection cable 15 m long stra...

Page 30: ...ent 41 Instrument settings 43 Interface commands 49 Interface connection 38 Internal temperature of the pyrometer 46 L Laser targeting light 36 41 M Maintenance 54 Maximum value storage 37 45 Measurem...

Page 31: ...IMPAC Infrared GmbH Temperaturmessger te Kleyerstra e 90 D 60326 Frankfurt Main Tel 49 0 69 9 73 73 0 Fax 49 0 69 9 73 73 167 Internet www impacinfrared com E Mail info impacinfrared com 3 869 207 a...

Reviews: