0.01...25 s: 

If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in 

double storage mode.

 After the entered time the storage will be deleted. 

extern: 

The external clearing can be activated and used within an own software (see section 

14

,

 Data 

format UPP

®

) or via an external contact (for connection see 

4.1 Connector pin assignment 

on the back side of the pyrometer

). In this case, the storage operates only in 

single storage

because only a single deletion mechanism is used.  

auto: 

The “

auto

” mode is used for discontinuous measuring tasks. For example objects are trans-

ported on a conveyer belt and pass the measuring beam of the pyrometer only for a few sec-
onds. Here the maximum value for each object has to be indicated. In this mode the maximum 
value is stored until a new hot object appears in the measuring beam. The temperature which 
has to be recognized as “hot“ is defined by the low limit of the adjusted sub range. The stored 
maximum value will be deleted when the temperature of the new hot object exceeds the low 
limit “

from

“ of the sub range by 1% or at least 2°C. If a lower limit is not entered, the maximum 

value storage will be deleted whenever the lower level of the full measuring range has been ex-
ceeded.  

Hold:

 

The function “hold” enables to freeze the current temperature reading at any moment. For this 
an external push button or switch has to be connected (see connector pin J  in 

4.1

) which holds 

the temperature reading as long as the contacts are closed.  

 

Operation note:

 dependent on the settings the maximum value storage either works in 

single storage

 mode 

or in 

double storage 

mode: 

 

Single storage: 

the 

single storage

 is used when you want to reset the stored value using an external im-

pulse via 

one 

contact closure from an external relay (i.e. between two measured objects). The relay contact 

is connected directly to the pyrometer between pins J and K. This mode allows a new value to be estab-
lished, after each impulse from the reset signal. 

 

double storage: 

when entering the reset intervals via push buttons or PC interface the 

double storage

 is 

automatically selected. This mode utilizes 

two

 memories in which the highest measured value is held and is 

deleted alternately in the time interval set (clear time). The other memory retains the maximum value 
throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency 
are thereby eliminated. 
 

 

 

Note:  

It doesn’t make sense to use the maximum value storage if an integrated 

PID con-

troller

 is used. Therefore the clear time should be set to

 

OFF position. 

 The 

external delete 

function

 

cannot be used when the PID controller is active, as 

the switch contact (connector pin J) is used for emergency switch-off. If the maxi-
mum value storage is set to 

external delete

 when the PID controller is activated, 

the clear time automatically will be set to OFF. 

 

 

 

Note:  

The maximum value storage follows the function of adjustment of exposure time.  
This results in: 

•  

clear time 

 the adjusted exposure time is useless 

 

 

•  

clear times must be at least 3 times longer than the exposure time 

 

 

•  

only maxima with full maximum value can be recorded, which 

 

   appear at least 3 times longer than the exposure time. 

 
 

9.2 

Internal temperature of the pyrometer 

The internal temperature of the pyrometer can be read on the interface. It is a few degrees higher than the 
ambient temperature due to the heat generated by the electronics.  
 
 

9.3 Bus 

address 

For the connecting of several pyrometers with RS485 with one serial interface it is 
necessary to give each instrument an individual address for communication. First it is 
necessary to connect each single instrument to give it an address. After that all instru-
ments can be connected and addressed individually. If parameters may be changed 
simultaneously on all pyrometers, the global 

Address 98

 can be used. This allows you to program all py-

rometers at the same time, regardless of the addresses that have already been assigned. If the address of a 
pyrometer is unknown, it is possible to communicate with it using the global 

Address

 

99

 (connect only one 

pyrometer). 

Settings: 

00 

. . . 

97 

Summary of Contents for Impac IGA 5

Page 1: ...IS 5 IGA 5 IS 5 LO IGA 5 LO INFRATHERM Pyrometer IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry Betriebsanleitung User Manual...

Page 2: ...S485 38 4 5 Connection schematic for analysing devices 39 5 Mechanical installation 39 5 1 Accessories option 39 5 2 Connection overview IS 5 LO IGA 5 LO 40 6 Instrument alignment 41 6 1 View finder 4...

Page 3: ...48 10 8 Configuration of the display on the TV screen 49 10 8 1 The PID controller 50 10 8 2 Self tuning algorithm 51 10 9 Measurement color bar 51 10 10 Measurement online trend 52 10 11 Output listi...

Page 4: ...otify any person or persons of such revisions or changes All series 5 Instruments from IMPAC Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects...

Page 5: ...isolated from each other Parameters Adjustable on the pyrometer Emissivity exposure time t90 analog output 0 20 or 4 20 mA online offline switch for adjustments at pyrometer or PC Additionally adjusta...

Page 6: ...levels controlled by the measuring temperature Field of view approx 10 x 14 of focused distance Date time real time clock with at least 3 days spare run Video output plug separate round plug at the p...

Page 7: ...if the person already used similar instruments or was already trained by the manu facturer The pyrometer has only to be used for the purpose described in the manual It is recommended to use only acces...

Page 8: ...e via external switch J pink see 4 1 1 external clearing of max value storage or hold function Only pyrometers with PID controller controller activated set controller output to 0 emergency stop G red...

Page 9: ...r fast measuring equipment So it is recommended to use the IMPAC converter order no 3 852 430 4 3 Connection via interface RS232 The transmission rate in baud of the serial interface is dependent on t...

Page 10: ...made from stainless steel protects the pyrometer if exposed to a hot environment It is designed for ambient temperatures up to 180 C Water cooling jacket with integrated air purge Miscellaneous for IS...

Page 11: ...way to align the pyrometer to the measured object The clamping screws of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions Mounting angle Ball and sock...

Page 12: ...Attention Do not look directly into the laser beam Laser class 2 according to IEC 60825 1 3 4 Note The laser warning signs on the pyrometer should be easily viewable at all times even after it has bee...

Page 13: ...2 15 7 2 Differing to the focused measuring distance all instrument types The optical profiles show that the spot size M varies with the distance from the pyrometer to the measuring object Spot sizes...

Page 14: ...out bending or twisting it Caution Disconnect the cable before opening the cover When reassembling the cover insert it carefully into the guide pin and then fasten it with the screws After that the co...

Page 15: ...r a correct measurement it is necessary to adjust the emissivity This emissivity is the relationship between the emission of an real object and the emission of a black body radiation source this is an...

Page 16: ...ttings The digital PC interface enables the exchange of data with a PC either by using the InfraWin software sup plied or an user adapted communication software see command table in section 14 Data fo...

Page 17: ...stab lished after each impulse from the reset signal double storage when entering the reset intervals via push buttons or PC interface the double storage is automatically selected This mode utilizes t...

Page 18: ...n of the individual icons in the program s help menu Click on the F1 button after InfraWin has been loaded or click on the in the menu bar The following descriptions refer to program version 4 0 The l...

Page 19: ...of devices InfraWin changes to the display of 1 or 2 devices If 2 devices are selected always 2 windows are displayed for settings or evaluation 10 7 Basic settings Under pyrometer parameters all pre...

Page 20: ...indow the connection with the preset baud rate can be checked Here the command was send 500 times with 19200 baud It has taken 4 56 seconds and no transmission errors has occurred 10 8 Configuration o...

Page 21: ...sired temperature a corresponding signal which changes over the time is transmitted to the analog output until the deviation reaches zero The integral time can be set in two different formats min s fo...

Page 22: ...red temperature Problems are generally posed by directly controlled systems where there are high contrasts between the cooling time constant and the heating up time constant For example in many cases...

Page 23: ...cally saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old values are over written when a new measurement is taken Files from older pr...

Page 24: ...er file the previous file will be overwritten and replaced by the standard i12 file 10 14 PC sampling rate time interval between two measurements This function sets a time interval After each interval...

Page 25: ...lower correct emissivity corresponding to the material see 8 5 Lens contaminated Clean lens carefully see 12 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corre...

Page 26: ...repeated AAmsXXX XXX 000 999 auto repeat of the ms command Reading basic temperature range AAmb Output XXXXYYYY 2 x 4 hexdigit for MBA and MBE C or F Reading sub range AAme Output XXXXYYYY 2 x 4 hexdi...

Page 27: ...rtional control Reset time Integral time AATiXX XX AATiXX XX or XX XX Ti time in format min sec XX XX is equivalent to 00 01 99 99 calculation is every second Ti time in Format XX XX sec is equivalent...

Page 28: ...r RS485 IS 5 MB 20 600 2000 C 3 857 150 3 857 170 3 857 160 3 857 180 IS 5 MB 25 800 2500 C 3 857 200 3 857 220 3 857 210 3 857 230 IS 5 MB 30 1000 3000 C 3 857 250 3 857 270 3 857 260 3 857 280 IGA 5...

Page 29: ...justable for optical head I 3 820 330 Connection cable 5 m long straight pyrometer connector 3 820 500 Connection cable 10 m long straight pyrometer connector 3 820 510 Connection cable 15 m long stra...

Page 30: ...ent 41 Instrument settings 43 Interface commands 49 Interface connection 38 Internal temperature of the pyrometer 46 L Laser targeting light 36 41 M Maintenance 54 Maximum value storage 37 45 Measurem...

Page 31: ...IMPAC Infrared GmbH Temperaturmessger te Kleyerstra e 90 D 60326 Frankfurt Main Tel 49 0 69 9 73 73 0 Fax 49 0 69 9 73 73 167 Internet www impacinfrared com E Mail info impacinfrared com 3 869 207 a...

Reviews: