Model 3001 Multi-Detector Survey Meter
User’s Manual
Section 4
Ludlum Measurements, Inc.
Page 4-13
April 2017
Setup Page 3
Secondary RATE/MAX Units and Minimum Display (Default 000
cpm
)
-
Tap ON/ACK to select the value to adjust, and MODE to adjust the value.
Available values are:
Number of Decimal Places (0, 1, or 2)
Minimum Units (See list below.)
cpm kcpm
cps
kcps
dpm
kdpm
Bq
OFF
kBq µR/h mR/h R/h µSv/h mSv/h Sv/h
Secondary Units RATE/MAX Maximum Display (Default 999 kcpm) -
Tap ON/ACK to select the value to adjust, and MODE to adjust the value. If
the Secondary Units is off, this parameter will be skipped. Units will be the same
as selected earlier with Secondary Units. Available values are:
Ones Place (
0-9
)
Tens Place (
0-9
)
Hundreds Place (
0-9
)
Number of Decimal Places (0, 1, or 2)
Range (
k
on or off – cpm, cps, dpm, Bq; µ, m or none for R/h and Sv/h)
Secondary Units RATE/MAX Mode Alarm Point (Default 000) -
Tap
ON/ACK to select the value to adjust, and MODE to adjust the value. If the
Secondary Units is off, this parameter will be skipped. Otherwise, units will be
the same as selected earlier with Secondary Units. The ALARM LCD Segment
will be on to indicate an Alarm parameter. Set this Alarm Point to 000 to
disable. Available values are:
Ones Place (
0-9
)
Tens Place (
0-9
)
Hundreds Place (
0-9
)
Number of Decimal Places (0, 1, or 2)
Range (
k
on or off – cpm, cps, dpm, Bq; µ, m or none for R/h and Sv/h)
Note:
If the Secondary Units has changed to a value other than that used to
previously set this Alarm Point, the Alarm Point will be reset to 000.