Faults
MXD70 Auxiliary mA Input
Setup and Operating Guide
- 35 -
Faults
E041 CH1 Partial
Depletion
E091 CH2
Large difference between the detectors, i.e. one very dirty. This alarm will come up
if there is a difference of 3:1 between the detectors. Remove sensor and clean
sensor fingers. The probe sensor units (PSU) will be set to 16000 or 32000 if
turbidity. If this message persists, please consult with your supplier.
E141 CH3
E042 CH1 Full
Depletion
E092 CH2
Attenuation too high or the real probe signal goes above 14000. The probe sensor
units (PSU) will be set to 16000. If this message persists, please consult with your
supplier.
E142 CH3
E043
CH1
Sensor User Offset At Limit
E093 CH2
The last Sensor Offset Calibration was out of limits, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
E143 CH3
E044
CH1
Sensor User Slope At Limit
E094 CH2
The last Sensor Slope Calibration was out of limits, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
E144 CH3
E045
CH1
Sensor User Slope < Spec
E095 CH2
The last Sensor Slope Calibration was less than the recommended specification,
check sensor condition and connections and repeat calibration. If the message
persists please consult with your supplier.
E145 CH3
E046
CH1
Sensor User Slope > Spec
E096 CH2
The last Sensor Slope Calibration was greater than the recommended specification,
check sensor condition and connections and repeat calibration. If the message
persists please consult with your supplier.
E146 CH3
E047
CH1
Sensor Open Circuit
E097 CH2
The sensor input is at open circuit, check sensor condition and connections. If the
message persists please consult with your supplier.
E147 CH3
E048
CH1
Sensor Short Circuit
E098 CH2
The sensor input is at short circuit, check sensor condition and connections. If the
message persists please consult with your supplier.
E148 CH3
E049
CH1
Sensor Positive Saturation
E099 CH2
The sensor input is greater than the maximum measurable input level, Check
Sensor condition and connections. If the message persists please consult with your
supplier.
E149 CH3
E050 CH1 Sensor
Negative
Saturation
E100 CH2
The sensor input is less than the minimum measurable input level, Check Sensor
condition and connections. If the message persists please consult with your
supplier.
E150 CH3
E051
CH1
Sensor Input Over Range
E101 CH2
The sensor reading is greater than the specified upper limit, check channel settings,
Sensor condition and connections. If the message persists please consult with your
supplier.
E151 CH3
E052
CH1
Sensor Input Under Range
E102 CH2
The sensor reading is less than the specified limit, check channel settings, Sensor
condition and connections. If the message persists please consult with your
supplier.
E152 CH3
Summary of Contents for MXD70 Series
Page 6: ...MXD73 Multi parameter Monitor Installation Guide ...
Page 7: ......
Page 10: ...Preface MXD73 Installation Guide 3 Declaration of Conformity ...
Page 11: ...Preface 4 MXD73 Installation Guide Blank ...
Page 13: ...Contents 6 MXD73 Installation Guide Blank ...
Page 15: ...Introduction 8 MXD73 Installation Guide Introduction MXD73 Rear Termination Overview ...
Page 25: ...Installation 18 MXD73 Installation Guide Installation Blank ...
Page 33: ...Notes 26 MXD73 Installation Guide ...
Page 34: ......
Page 36: ...MXD75 Multi parameter Monitor Installation Guide ...
Page 37: ......
Page 40: ...Preface MXD75 Installation Guide 3 Declaration of Conformity ...
Page 41: ...Preface 4 MXD75 Installation Guide Blank ...
Page 43: ...Contents 6 MXD75 Installation Guide Blank ...
Page 45: ...Introduction 8 MXD75 Installation Guide Introduction MXD75 Termination Overview ...
Page 58: ...Installation MXD75 Installation Guide 21 Installation Horizontal Mounting Vertical Mounting ...
Page 59: ...Installation 22 MXD75 Installation Guide Installation Blank ...
Page 67: ...Notes 30 MXD75 Installation Guide ...
Page 68: ...Notes MXD75 Installation Guide 31 ...
Page 70: ......
Page 72: ...MXD70 SERIES Multi parameter Monitor User Interface Data Logging Guide ...
Page 73: ......
Page 77: ...Contents 4 User Interface Guide Blank Page ...
Page 81: ...User Interface 8 User Interface Guide User Interface Blank Page ...
Page 85: ...Security Code Access 12 User Interface Guide Security Code Access Blank Page ...
Page 93: ...Update Software 20 User Interface Guide Update Software Blank Page ...
Page 107: ...Data Logging 34 User Interface Guide Data Logging Blank Page ...
Page 109: ...Error Messages 36 User Interface Guide Error Messages Blank Page ...
Page 116: ......
Page 117: ......
Page 118: ...MXD70 SERIES Multi parameter Monitor Conductivity Setup and Operating Guide ...
Page 119: ......
Page 123: ...Contents 4 MXD70 Conductivity Setup and Operating Guide Blank ...
Page 133: ...Installation 14 MXD70 Conductivity Setup and Operating Guide Installation Blank ...
Page 149: ...Channel Calibration 30 MXD70 Conductivity Setup and Operating Guide Channel Calibration Blank ...
Page 155: ...Appendix 36 MXD70 Conductivity Setup and Operating Guide Appendix ...
Page 177: ...Notes 58 MXD70 Conductivity Setup and Operating Guide ...
Page 178: ...Notes MXD70 Conductivity Setup and Operating Guide 59 ...
Page 180: ......
Page 181: ......
Page 182: ...MXD70 SERIES Multi parameter Monitor pH Redox Setup and Operating Guide ...
Page 183: ......
Page 211: ...Channel Setup 28 MXD70 pH Redox Setup and Operating Guide Channel Setup Blank ...
Page 251: ...Notes 68 MXD70 pH Redox Setup and Operating Guide ...
Page 252: ...Notes MXD70 pH Redox Setup and Operating Guide 69 ...
Page 254: ......
Page 256: ...MXD70 SERIES Multi parameter Monitor Electrodeless Conductivity Setup and Operating Guide ...
Page 257: ......
Page 261: ...Contents 4 MXD70 Electrodeless Conductivity Setup and Operating Guide Blank ...
Page 295: ......
Page 311: ...Notes 54 MXD70 Electrodeless Conductivity Setup and Operating Guide ...
Page 313: ......
Page 314: ......
Page 315: ...MXD70 SERIES Multi parameter Monitor Dissolved Oxygen Setup and Operating Guide ...
Page 316: ......
Page 320: ...Contents 4 MXD70 Dissolved Oxygen Setup and Operating Guide Blank ...
Page 322: ...Specification 6 MXD70 Dissolved Oxygen Setup and Operating Guide Specification Blank ...
Page 339: ...Channel Setup MXD70 Dissolved Oxygen Setup and Operating Guide 23 Channel Setup Blank ...
Page 353: ...Channel Calibration MXD70 Dissolved Oxygen Setup and Operating Guide 37 Channel Calibration ...
Page 359: ...Appendix MXD70 Dissolved Oxygen Setup and Operating Guide 43 Appendix Blank ...
Page 381: ...Notes MXD70 Dissolved Oxygen Setup and Operating Guide 65 ...
Page 382: ...Notes 66 MXD70 Dissolved Oxygen Setup and Operating Guide ...
Page 384: ......
Page 385: ......
Page 386: ...MXD70 SERIES Multi parameter Monitor Auxiliary mA Input Setup and Operating Guide ...
Page 387: ......
Page 391: ...Contents 4 MXD70 Auxiliary mA Input Setup and Operating Guide Blank ...
Page 393: ...Specification 6 MXD70 Auxiliary mA Input Setup and Operating Guide Specification Blank ...
Page 405: ...Channel Setup 18 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Setup BLANK ...
Page 434: ...Notes MXD70 Auxiliary mA Input Setup and Operating Guide 47 ...
Page 436: ......
Page 437: ......
Page 438: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Page 439: ......
Page 443: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Page 465: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Page 486: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Page 488: ......
Page 489: ......
Page 491: ......
Page 495: ...Contents 4 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Blank ...
Page 511: ...Setpoints 20 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Setpoints ...
Page 525: ...Notes 34 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide ...
Page 526: ...Notes MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide 35 ...
Page 528: ......
Page 529: ......
Page 530: ...MXD70 Series Multi parameter Monitor Modbus RS485 Interface Operating Guide ...
Page 531: ......
Page 535: ...Contents 4 Modbus RS485 Interface Operating Guide BLANK ...
Page 539: ...Modbus RS485 8 Modbus RS485 Interface Operating Guide Modbus RS485 BLANK ...
Page 628: ...Notes Modbus RS485 Interface Operating Guide 97 ...
Page 629: ...Notes 98 Modbus RS485 Interface Operating Guide ...
Page 630: ......