Faults
- 40 -
MXD70 Electrodeless Conductivity
Setup and Operating Guide
Faults
Input Channel Errors
E030
CH1
Input Card Checksum Error
E080
CH2
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as the channel’s input card may require to be returned for repair.
E130
CH3
E031 CH1 Setup
Checksum
Error
E081 CH2
The current channel’s configuration has for some reason become corrupted. Try
switching the unit off and then on again. If the message persists reset the current
channel from the Reset Unit Settings option in the Save/Restore menu or consult
with your supplier, as the channel’s input card may require to be returned for
repair.
E131 CH3
E032
CH1
Store A Checksum Error
E082 CH2
The data in the channel’s Store A has become corrupted. Check the channel’s
current setup. Then save the setup back to channel’s Store A in the Save/Restore
menu.
E132 CH3
E033
CH1
Store B Checksum Error
E083 CH2
The data in the channel’s Store B has become corrupted. Check the channel’s
current setup. Then save the setup back to channel’s Store B in the Save/Restore
menu.
E133 CH3
E034
CH1
Factory Cal Checksum Error
E084 CH2
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as the channel’s input card may require to be returned for repair.
E134 CH3
E035
CH1
User Cal Checksum Error
E085 CH2
The Channel’s User Cal has for some reason become corrupted. Try switching the
unit off and then on again. If the message persists Reset the Channel from the
Reset User Calibration option in the Calibration menu or consult with your supplier,
as the channel’s input card may require to be returned for repair.
E135 CH3
E036
CH1
Sensor Cal Out Of Spec
E086 CH2
The last Sensor Calibration was out of specification, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
E136 CH3
E037
CH1
Sensor Zero Cal Out Of Spec
E087 CH2
The last Sensor Zero Calibration was out of specification, check sensor condition
and connections and repeat calibration. If the message persists please consult with
your supplier.
E137 CH3
E038
CH1
Sensor Span Cal Out Of Spec
E088 CH2
The last Sensor Span Calibration was out of specification, check sensor condition
and connections and repeat calibration. If the message persists please consult with
your supplier.
E138 CH3
E039 CH1 No
Signal
E089 CH2
No Sensor connected or there is 'zero' detector current. The probe sensor units
(PSU) will be set to 16000 or 32000 if turbidity. If this message persists, please
consult with your supplier.
E139 CH3
E040 CH1 Signal
Overload
E090 CH2
The gain step is equal to 0 and the A/D output is over 255. This cannot happen in a
liquid but could happen if the sensor is in full sunlight. The probe sensor units
(PSU) will be set to 16000 or 32000 if turbidity. If this message persists, please
consult with your supplier.
E140 CH3
Summary of Contents for MXD70 Series
Page 6: ...MXD73 Multi parameter Monitor Installation Guide ...
Page 7: ......
Page 10: ...Preface MXD73 Installation Guide 3 Declaration of Conformity ...
Page 11: ...Preface 4 MXD73 Installation Guide Blank ...
Page 13: ...Contents 6 MXD73 Installation Guide Blank ...
Page 15: ...Introduction 8 MXD73 Installation Guide Introduction MXD73 Rear Termination Overview ...
Page 25: ...Installation 18 MXD73 Installation Guide Installation Blank ...
Page 33: ...Notes 26 MXD73 Installation Guide ...
Page 34: ......
Page 36: ...MXD75 Multi parameter Monitor Installation Guide ...
Page 37: ......
Page 40: ...Preface MXD75 Installation Guide 3 Declaration of Conformity ...
Page 41: ...Preface 4 MXD75 Installation Guide Blank ...
Page 43: ...Contents 6 MXD75 Installation Guide Blank ...
Page 45: ...Introduction 8 MXD75 Installation Guide Introduction MXD75 Termination Overview ...
Page 58: ...Installation MXD75 Installation Guide 21 Installation Horizontal Mounting Vertical Mounting ...
Page 59: ...Installation 22 MXD75 Installation Guide Installation Blank ...
Page 67: ...Notes 30 MXD75 Installation Guide ...
Page 68: ...Notes MXD75 Installation Guide 31 ...
Page 70: ......
Page 72: ...MXD70 SERIES Multi parameter Monitor User Interface Data Logging Guide ...
Page 73: ......
Page 77: ...Contents 4 User Interface Guide Blank Page ...
Page 81: ...User Interface 8 User Interface Guide User Interface Blank Page ...
Page 85: ...Security Code Access 12 User Interface Guide Security Code Access Blank Page ...
Page 93: ...Update Software 20 User Interface Guide Update Software Blank Page ...
Page 107: ...Data Logging 34 User Interface Guide Data Logging Blank Page ...
Page 109: ...Error Messages 36 User Interface Guide Error Messages Blank Page ...
Page 116: ......
Page 117: ......
Page 118: ...MXD70 SERIES Multi parameter Monitor Conductivity Setup and Operating Guide ...
Page 119: ......
Page 123: ...Contents 4 MXD70 Conductivity Setup and Operating Guide Blank ...
Page 133: ...Installation 14 MXD70 Conductivity Setup and Operating Guide Installation Blank ...
Page 149: ...Channel Calibration 30 MXD70 Conductivity Setup and Operating Guide Channel Calibration Blank ...
Page 155: ...Appendix 36 MXD70 Conductivity Setup and Operating Guide Appendix ...
Page 177: ...Notes 58 MXD70 Conductivity Setup and Operating Guide ...
Page 178: ...Notes MXD70 Conductivity Setup and Operating Guide 59 ...
Page 180: ......
Page 181: ......
Page 182: ...MXD70 SERIES Multi parameter Monitor pH Redox Setup and Operating Guide ...
Page 183: ......
Page 211: ...Channel Setup 28 MXD70 pH Redox Setup and Operating Guide Channel Setup Blank ...
Page 251: ...Notes 68 MXD70 pH Redox Setup and Operating Guide ...
Page 252: ...Notes MXD70 pH Redox Setup and Operating Guide 69 ...
Page 254: ......
Page 256: ...MXD70 SERIES Multi parameter Monitor Electrodeless Conductivity Setup and Operating Guide ...
Page 257: ......
Page 261: ...Contents 4 MXD70 Electrodeless Conductivity Setup and Operating Guide Blank ...
Page 295: ......
Page 311: ...Notes 54 MXD70 Electrodeless Conductivity Setup and Operating Guide ...
Page 313: ......
Page 314: ......
Page 315: ...MXD70 SERIES Multi parameter Monitor Dissolved Oxygen Setup and Operating Guide ...
Page 316: ......
Page 320: ...Contents 4 MXD70 Dissolved Oxygen Setup and Operating Guide Blank ...
Page 322: ...Specification 6 MXD70 Dissolved Oxygen Setup and Operating Guide Specification Blank ...
Page 339: ...Channel Setup MXD70 Dissolved Oxygen Setup and Operating Guide 23 Channel Setup Blank ...
Page 353: ...Channel Calibration MXD70 Dissolved Oxygen Setup and Operating Guide 37 Channel Calibration ...
Page 359: ...Appendix MXD70 Dissolved Oxygen Setup and Operating Guide 43 Appendix Blank ...
Page 381: ...Notes MXD70 Dissolved Oxygen Setup and Operating Guide 65 ...
Page 382: ...Notes 66 MXD70 Dissolved Oxygen Setup and Operating Guide ...
Page 384: ......
Page 385: ......
Page 386: ...MXD70 SERIES Multi parameter Monitor Auxiliary mA Input Setup and Operating Guide ...
Page 387: ......
Page 391: ...Contents 4 MXD70 Auxiliary mA Input Setup and Operating Guide Blank ...
Page 393: ...Specification 6 MXD70 Auxiliary mA Input Setup and Operating Guide Specification Blank ...
Page 405: ...Channel Setup 18 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Setup BLANK ...
Page 434: ...Notes MXD70 Auxiliary mA Input Setup and Operating Guide 47 ...
Page 436: ......
Page 437: ......
Page 438: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Page 439: ......
Page 443: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Page 465: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Page 486: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Page 488: ......
Page 489: ......
Page 491: ......
Page 495: ...Contents 4 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Blank ...
Page 511: ...Setpoints 20 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Setpoints ...
Page 525: ...Notes 34 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide ...
Page 526: ...Notes MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide 35 ...
Page 528: ......
Page 529: ......
Page 530: ...MXD70 Series Multi parameter Monitor Modbus RS485 Interface Operating Guide ...
Page 531: ......
Page 535: ...Contents 4 Modbus RS485 Interface Operating Guide BLANK ...
Page 539: ...Modbus RS485 8 Modbus RS485 Interface Operating Guide Modbus RS485 BLANK ...
Page 628: ...Notes Modbus RS485 Interface Operating Guide 97 ...
Page 629: ...Notes 98 Modbus RS485 Interface Operating Guide ...
Page 630: ......