Channel Calibration
MXD70 pH / Redox
Setup and Operating Guide
- 41 -
Channel Cal
ibration
Manual pH Sensor Calibration
When using Manual calibration mode it is possible to do either a single point Buffer (Offset) Cal or a
combined two point Buffer and Slope calibration. This is accomplished by using either the
pH Buffer
(Offset) Cal
or
pH Slope Cal
menu items. The combined Buffer and Slope calibration is done in
the following order:
1. Adjustment of the Buffer (Offset) value at 7.00 pH.
2. Adjustment of the Slope (Gain) value at any other pH (usually 4 or 9 pH).
As actual pH buffer values are used no compensation is made for the variation of the buffer solution
with temperature in the instrument. It is important therefore to know the actual buffer value at the
temperature of the solution. The standard LTH 4, 7 and 9 pH values at temperature can be seen on
page 44.
It is also important that the buffer temperature is entered into the Calibration Manual Temperature
Input if manual temperature compensation is being employed. Alternatively any automatic
temperature compensation element should be placed in the buffer solution with the pH sensor if Auto
TC is being used.
Calibration Principle
The default method of calibration is “Manual” mode,
however if the probe is pH and the type is glass then
“Auto” mode may have been selected. If this is the
case then calibration principle will need to be set to
manual.
/
– Select Option
EXIT
– Cancel
– Save Selection
pH Slope Calibration – Buffer Calibration
The pH buffer calibration enables the user to adjust
the sensor buffer until the displayed reading matches
the known input. To activate the function, select “pH
Slope Cal” and press enter.
The current sensor reading can be seen in the pop-up
window and is adjusted by pressing the up and down
arrows. When the reading is correct press the enter
button to store the calibration and progress to the
slope calibration.
It is recommended that the buffer should be adjusted
at 7pH (
±
2pH) for a glass probe, and 0pH (+2pH) for
an antimony probe.
Note. If the electrode is new, pressing the “new”
button will reset the existing offset.
/
– Adjust the Reading Up or Down
NEW
– Register New Electrode
EXIT
– Cancel
– Progress to Slope Calibration
Summary of Contents for MXD70 Series
Page 6: ...MXD73 Multi parameter Monitor Installation Guide ...
Page 7: ......
Page 10: ...Preface MXD73 Installation Guide 3 Declaration of Conformity ...
Page 11: ...Preface 4 MXD73 Installation Guide Blank ...
Page 13: ...Contents 6 MXD73 Installation Guide Blank ...
Page 15: ...Introduction 8 MXD73 Installation Guide Introduction MXD73 Rear Termination Overview ...
Page 25: ...Installation 18 MXD73 Installation Guide Installation Blank ...
Page 33: ...Notes 26 MXD73 Installation Guide ...
Page 34: ......
Page 36: ...MXD75 Multi parameter Monitor Installation Guide ...
Page 37: ......
Page 40: ...Preface MXD75 Installation Guide 3 Declaration of Conformity ...
Page 41: ...Preface 4 MXD75 Installation Guide Blank ...
Page 43: ...Contents 6 MXD75 Installation Guide Blank ...
Page 45: ...Introduction 8 MXD75 Installation Guide Introduction MXD75 Termination Overview ...
Page 58: ...Installation MXD75 Installation Guide 21 Installation Horizontal Mounting Vertical Mounting ...
Page 59: ...Installation 22 MXD75 Installation Guide Installation Blank ...
Page 67: ...Notes 30 MXD75 Installation Guide ...
Page 68: ...Notes MXD75 Installation Guide 31 ...
Page 70: ......
Page 72: ...MXD70 SERIES Multi parameter Monitor User Interface Data Logging Guide ...
Page 73: ......
Page 77: ...Contents 4 User Interface Guide Blank Page ...
Page 81: ...User Interface 8 User Interface Guide User Interface Blank Page ...
Page 85: ...Security Code Access 12 User Interface Guide Security Code Access Blank Page ...
Page 93: ...Update Software 20 User Interface Guide Update Software Blank Page ...
Page 107: ...Data Logging 34 User Interface Guide Data Logging Blank Page ...
Page 109: ...Error Messages 36 User Interface Guide Error Messages Blank Page ...
Page 116: ......
Page 117: ......
Page 118: ...MXD70 SERIES Multi parameter Monitor Conductivity Setup and Operating Guide ...
Page 119: ......
Page 123: ...Contents 4 MXD70 Conductivity Setup and Operating Guide Blank ...
Page 133: ...Installation 14 MXD70 Conductivity Setup and Operating Guide Installation Blank ...
Page 149: ...Channel Calibration 30 MXD70 Conductivity Setup and Operating Guide Channel Calibration Blank ...
Page 155: ...Appendix 36 MXD70 Conductivity Setup and Operating Guide Appendix ...
Page 177: ...Notes 58 MXD70 Conductivity Setup and Operating Guide ...
Page 178: ...Notes MXD70 Conductivity Setup and Operating Guide 59 ...
Page 180: ......
Page 181: ......
Page 182: ...MXD70 SERIES Multi parameter Monitor pH Redox Setup and Operating Guide ...
Page 183: ......
Page 211: ...Channel Setup 28 MXD70 pH Redox Setup and Operating Guide Channel Setup Blank ...
Page 251: ...Notes 68 MXD70 pH Redox Setup and Operating Guide ...
Page 252: ...Notes MXD70 pH Redox Setup and Operating Guide 69 ...
Page 254: ......
Page 256: ...MXD70 SERIES Multi parameter Monitor Electrodeless Conductivity Setup and Operating Guide ...
Page 257: ......
Page 261: ...Contents 4 MXD70 Electrodeless Conductivity Setup and Operating Guide Blank ...
Page 295: ......
Page 311: ...Notes 54 MXD70 Electrodeless Conductivity Setup and Operating Guide ...
Page 313: ......
Page 314: ......
Page 315: ...MXD70 SERIES Multi parameter Monitor Dissolved Oxygen Setup and Operating Guide ...
Page 316: ......
Page 320: ...Contents 4 MXD70 Dissolved Oxygen Setup and Operating Guide Blank ...
Page 322: ...Specification 6 MXD70 Dissolved Oxygen Setup and Operating Guide Specification Blank ...
Page 339: ...Channel Setup MXD70 Dissolved Oxygen Setup and Operating Guide 23 Channel Setup Blank ...
Page 353: ...Channel Calibration MXD70 Dissolved Oxygen Setup and Operating Guide 37 Channel Calibration ...
Page 359: ...Appendix MXD70 Dissolved Oxygen Setup and Operating Guide 43 Appendix Blank ...
Page 381: ...Notes MXD70 Dissolved Oxygen Setup and Operating Guide 65 ...
Page 382: ...Notes 66 MXD70 Dissolved Oxygen Setup and Operating Guide ...
Page 384: ......
Page 385: ......
Page 386: ...MXD70 SERIES Multi parameter Monitor Auxiliary mA Input Setup and Operating Guide ...
Page 387: ......
Page 391: ...Contents 4 MXD70 Auxiliary mA Input Setup and Operating Guide Blank ...
Page 393: ...Specification 6 MXD70 Auxiliary mA Input Setup and Operating Guide Specification Blank ...
Page 405: ...Channel Setup 18 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Setup BLANK ...
Page 434: ...Notes MXD70 Auxiliary mA Input Setup and Operating Guide 47 ...
Page 436: ......
Page 437: ......
Page 438: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Page 439: ......
Page 443: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Page 465: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Page 486: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Page 488: ......
Page 489: ......
Page 491: ......
Page 495: ...Contents 4 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Blank ...
Page 511: ...Setpoints 20 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Setpoints ...
Page 525: ...Notes 34 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide ...
Page 526: ...Notes MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide 35 ...
Page 528: ......
Page 529: ......
Page 530: ...MXD70 Series Multi parameter Monitor Modbus RS485 Interface Operating Guide ...
Page 531: ......
Page 535: ...Contents 4 Modbus RS485 Interface Operating Guide BLANK ...
Page 539: ...Modbus RS485 8 Modbus RS485 Interface Operating Guide Modbus RS485 BLANK ...
Page 628: ...Notes Modbus RS485 Interface Operating Guide 97 ...
Page 629: ...Notes 98 Modbus RS485 Interface Operating Guide ...
Page 630: ......