Description of device
23
EN
Depending on the sensor attached, the signals below may be recorded and displayed.
•
PC mode:
the device automatically switches to PC mode as soon as the device is
connected to a PC via a USB cable. The analysis results for the recording currently saved
can be imported into the PC software. The device switches off after 5 minutes if no data
transfer takes place. The following settings can be programmed using the PC software
(from SOMNO
lab
Version 2.11):
– start time for the recording
– display language
– patient name
For software updates, contact your local specialist dealer or the manufacturer.
Diagnostic nasal cannula
The diagnostic nasal cannula detects respiratory flow and
snoring in combination with the pressure sensor
integrated in the device. Inspiration is registered by the
vacuum generated, exhalation by the overpressure
generated. Snoring generates pressure fluctuations in
the nostrils which are likewise registered.
Signal display and automatic analysis are matched to the
genuine diagnostic nasal cannula. SOMNO
check
micro
may only be used in conjunction with the genuine
diagnostic nasal cannula.
Pulsoximetry sensor / CARDIO-Sensor
The pulsoximetry sensor detects pulsoximetric signals, in
other words, the patient’s blood oxygen saturation and
pulse frequency. Its primary components are two LEDs
and a receiver diode.
Several oxygen saturation values are determined for each
pulse wave (split pulse wave algorithm). If a signal is
Sensor
Measured values
Pulsoximetry sensor
– Oxygen saturation
– Pulse frequency
Diagnostic nasal cannula
– Respiratory flow