36
Operation
EN
4. Ask your patient to put on the device and the sensors
him or herself under your supervision.
5. Show your patient how to change the batteries or
rechargeable batteries (see “5.7 Insert batteries” on
page 45 and “5.8 Use rechargeable batteries” on
page 46).
6. Show your patient how to perform a recording (see
“5.3 Perform a recording” on page 36).
7. Give your patient a set of new batteries or charged
rechargeable batteries.
8. Stow the sleep apnea diagnosis set and the patient’s
instructions for use in the carrying bag.
9. Give your patient the carrying bag.
5.3 Perform a recording
Manual recording
1. Put on the device (see “4.1 Put on device” on
page 31).
2. Put on the diagnostic nasal cannula (see “4.2 Put on
diagnostic nasal cannula” on page 31).
3. Put on the pulsoximetry sensor (see “4.3 Attach
pulsoximetry sensor” on page 33).
4. Press the button on the device.
5. Wait until the start screen disappears.
6. Check whether there are any signals.
7. Press the button on the device for 3 seconds.
The device performs a signal test (
Signal test
) for a
maximum of 2 minutes.