Chapter 1: General Information
Safety
•
Automatic compression release (decompression) is disabled when a localization
paddle is installed.
•
C-arm Force Limit is reached. The downward motorized movement of the C-arm is
limited to a force of 154 N, +44 N (35 lb, +10 lb).
2.4
Radiation Safety
The unit’s radiation safety is designed to comply with all requirements of 21CFR Part 1020,
IEC 601-2-7, and IEC 601-1-3. The shield is rated for a 0.5 mm Pb (lead) equivalence. To
ensure radiological protection, restrict access to the unit in accordance with local
regulations.
The exposure duration is limited by the following normal conditions:
•
The manual mAs timer
•
The Automatic Exposure Control circuits
The exposure duration is also limited by the following abnormal conditions:
•
Premature release of the x-ray exposure switch
•
Exceeding the preset “back-up time”
•
The independent safety hardware back-up timer
•
Detection of a generator fault