6-36
P/N 9-500-0255
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
C-Arm Components—Remove and Replace
3.2.2
X-ray Tube
Removal of the x-ray
tube (Figure 6-25) will
require complete
calibration of the x-ray
system and automatic
exposure control (AEC)
system. Access the X-
ray Tube by removing
the upper and lower
tubehead covers.
1. Remove the tube-
head covers as per
Section 3.2.1.
2. Remove the Top
and Upper Rear C-
arm covers as per
Section 3.1.2.
3. Remove the high
voltage lead from
the rear of the x-
ray tube. The
connector is a 1/4-
turn quick
disconnect.
4. Unscrew the
filament connector
from the
receptacle on the
rear of the x-ray
tube.
5. Remove the 2
screws that secure
the rear of the tube
to the mounting
cradle on the
tubehead chassis.
6. Remove the 4
mounting screws
that secure the x-
ray tube to the front of the tubehead chassis. These screws are accessed from
below, through the beam limiting assembly. Lift the tube out of the unit.
7. Position the replacement tube in the tubehead so that the threaded holes in the
front of the x-ray tube align with the mounting holes in the tubehead chassis.
8. Loosely install the front 4 and the rear 2 mounting screws.
9. Tighten the 4 front mounting screws completely, then tighten the 2 screws on
the rear of the tube.
Figure 6-25: X-ray Tube—Removal