Appendix A: Specifications
System Specifications
for consecutive exposures, per 21CFR,
1020.31. Internal limit = 0.09.
kV Accuracy
The actual value will not differ by more than
+1 kV from the indicated value.
mAs Accuracy
±5% or ±2 mAs, whichever is greater, from
indicated, measured from the ground side of
the tube circuit.
Post mAs Accuracy
±5% or ±2 mAs, whichever is greater, from
actual mAs
mA Accuracy
The actual value will not differ by more than
+5% for the indicated value.
Table A-1: Kv versus mA table
kV
Large Focus
Large Focus
Reduced mA
Small Focus
20
75
20
20
21
80
22
22
22
85
24
24
23
90
26
26
24
95
28
28
25
100
30
30
26
100
30
30
27
100
30
30
28
100
30
30
29
100
30
30
30
100
30
30
31
100
30
30
32
100
30
30
33
85
28
28
34
80
28
28
35
80
26
26
36
75
26
26
37
75
26
26
38
70
24
24
39
70
24
24