6-10
P/N 9-500-0255
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
Gantry Components—Remove and Replace
2.4
Left-Side Gantry Components
Refer to Figure 6-1 for location of left-side gantry components throughout the
following subsections.
2.4.1
High Voltage Generator Assembly
The High Voltage Generator Assembly consists of the following
components:
•
HV Multiplier
•
HV Diode PCB Assembly (part of HV Multiplier)
•
HV Inverter Board
•
Capacitor Board (located on bottom of HV Inverter Board)
•
HV Control Filament Board
WARNING!
Always remove power before performing any
removal or replacement procedure.
Caution:
Always observe Electrostatic Discharge (ESD) precautions
when working with electronics and electronic components.
Note…
The HV Generator is assembled, calibrated, and complied at
the factory as an assembly and must be replaced as an
assembly.
The type of mounting plate used with the HV Generator Assembly
determines the remove and replace procedures used. Affinity
Mammography Systems with a Serial Number between xxxmmyy0001
and xxxmmyy0018 (xxx represents the system series designator; mmyy
represents the month and year of manufacture) require removal of the
HV Inverter Board and the HV Control Filament Board prior to
removing the HV Generator Assembly Mounting plate (page 11). For
those Affinity Mammography Systems with a Serial Number of
xxxmmyy0019 or above, the HV Generator Assembly can be removed
as a single item (page 14). The unit Serial Number can be found on a
nameplate located on the rear of the Gantry. Refer to Chapter 1, Figure
1-4, Label 1.