Copyright © Lojer Oy, 2014 | 05/2014
14
3.2 ON/StandBy (Power ON/StandBy)
Press the ON/StandBy button on the hand control (Figure 5 (1)) in order to use the table. During start-up
sequence the LEDs flash as a string and the table cannot be adjusted. The table is ready for use, when the
indicator LEDs for power (Figure 5 (2)), floor locks (Figure 5 (6/7)) and patient orientation are lit. In order to
save the batteries, the power is turned off after a period of time (programmable 1-240 min) if no functions are
used. All functions can be reactivated by pressing the ON/StandBy button. Turn the power by pressing the
ON/StandBy button until the LEDs turn off.
By turning the power off from the ON/StandBy button, all functions are shut down. Use the
button when immediate cancellation of movement is necessary.
3.3 Working position (Floor locks down)/Transport position (floor locks up)
The table adjusts automatically to working/transport position when the button is pressed. The
movement stops only by shutting down the power (pressing the ON/StandBy- button). If shut
down the power in order to stop the movement, put the power back on to make sure that the
floor locks are completely up or down.
In order to set the table in working position, press the button (Figure 5 (6)) on the hand control. The floor
locks descend. Indicator LEDs (green and yellow) start flashing in turn. Short signal can be heard and the
indicator LED for working position ( Figure 5 (6)) is lit when the floor locks are completely down and the table
is in working position.
In order to set the table on transport position, press the button approx. 1 second (Figure 5 (7)) on the hand
control. The floor locks get up. . Indicator LEDs (green and yellow) start flashing in turn. Short signal can be
heard and the indicator LED for transport position ( Figure 5 (7)) is lit when the floor locks are completely up
and the table is in transport position.
For safety reasons only height and trendeleburg adjustments are operational, when the table
is in transport position.
In case of electric malfunction, the table can be returned to transport position manually by pulling out the
emergency valve located in the main control panel (Figure 3).
Open the hatch and pull out the knob (Figure 8).
The table will immediately lower the wheels (transport position). Push the knob back for normal operation.
Figure 8: Floor lock emergency valve