Copyright © Lojer Oy, 2014 | 05/2014
15
3.4 Patient orientation
The patient is positioned on the table in normal orientation, when the head is faced towards the long end of
the lower frame (see Figure 2). The patient is positioned on reverse patient orientation when the head is
faced towards the short end of the lower frame. The indicator LEDs on the hand control indicated the chosen
patient orientation: reverse patient orientation (Figure 5 (9)) and normal patient orientation
(Figure 5 (10)). In
order to change the patient orientation, press the button (Figure 5 (8)) approx. 3 seconds. The chosen
orientation will remain until you press the patient orientation button again. Chosen orientation will remain
stored when the power is turned off.
For safety reasons you have to press the button approx. 3 seconds in order to activate the
function.
The buttons on the hand control always operate according to the chosen patient orientation.
3.5 Adjustments
Maximum adjustment range of the table is presented below (Figure 9).
SQUEEZING HAZARD! Make sure that nothing gets between the structure or under the
device during adjustment.
3.5.1 Height adjustment
Adjust the height of the table with buttons on the hand control (Figure 5 (16 & 17)).
3.5.2 Back section adjustment
Adjust the angle of the back section with the buttons on the hand control (Figure 5 (20 & 21)).
3.5.3 Leg section adjustment
Adjust the angle of the leg section with the buttons on the hand control (Figure 5 (18 & 19)).
3.5.4 Lateral tilt
Tilt the table (patient) right by pressing the button on the hand control (Figure 5 (15)). Tilt to left with the
button on the hand control (Figure 5 (14)). Maximum tilt 20 ° (Figure 9), when the table is horizontally. When
the table is in trendelenburg/antitrendelenburg-position the tilt angle is limited.
3.5.5 Trendelenburg/Antitrendelenburg
Adjust the table to Trendelenburg-position (head down) with the button on the hand control (Figure 5 (23)).
Adjust the table to Antitrendelenburg-position (leg down) with the button on the hand control (Figure 5 (22)).
Return to the horizontal position either by pressing the Zero-position button (Figure 5 (13)) or before
mentioned Trendelenburg/Antitrendelenburg buttons.
Figure 9: Adjustment range of the operating table