Operation instructions for Salli F1, F2, F3, H2 and H4 Hospital bed / Version 1.3
Copyright © Lojer Oy, 2015
11
IMPORTANT! Ensure that there is nobody or anything between the rails
when they are being adjusted. Ensure that you hear a click,
that the rail is locked in the upper position.
IMPORTANT! When a patient is being transported, lower the adjustments
of the sleeping platform and the bed to lowest positions and
lift the rails up to minimise a patient’s risk of falling.
IMPORTANT! Always use mattress size which has been defined by
manufacturer for combination of side rail / bed model.
IMPORTANT! A rail, which is in the middle or intermediate position will not
prevent a patient’s fall from the bed. Consequently, if a rail is
in the middle position, it does not meet the requirements of
standard IEC 60601-2-52 and the patient safety is
endangered.
Summary of Contents for SALLI F1
Page 1: ...SALLI Hospital beds F1 F2 F3 H2 H4 Operating instructions Version 1 3 04 09 2015...
Page 28: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 25...
Page 29: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 26...
Page 30: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 27...
Page 31: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 28...
Page 32: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 29...
Page 33: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 30...
Page 34: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 31...
Page 35: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 32...
Page 36: ...OperationinstructionsforSalliF1 F2 F3 H2andH4Hospitalbed Version1 3 Copyright Lojer Oy 2015 33...