48
www.logosbio.com
<Bright Field, Green, & Red Images of Unsettled Cells>
4.3.2 Insert the loaded slide into the slide port of the instrument, ensuring that the loaded chamber is
inserted first into the slide port. The instrument analyzes only the first inserted chamber. Gently insert
the counting slide to the end.
Note
: After inserting the slide, LUNA-
FL™ only reads the first chamber. To read the second chamber,
it must be taken out, rotated, and inserted again.
Note
: Make sure that the counting slide is not inserted upside-down. This may lead to sample spilling
and could severely damage the counter.
4.3.3 Select the “Fluorescence Cell Counting” button on the home screen.
<Home Screen>
Cells move in this direction.