![LOGIKODIGIT DM490 User Manual Download Page 45](http://html1.mh-extra.com/html/logikodigit/dm490/dm490_user-manual_1935629045.webp)
45
ENGLISH
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity -
The devices are intended for use in the electromagnetic environment specified below. The user
of this device should make sure it is used in such an environment.
Immunity Test
IEC60601
test level
Compliance
Electromagnetic
Environment-Guidance
Conducted RF
IEC 61000-4-6
3Vrms150
kHz to 80
MHz outside
ISM bandsa
3 Vrms
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of this device,
including cables, than there com-
mended separation distance calcu-
lated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 80 MHz to 2.5G MHz
Where P is the maximum output po-
wer rating of the transmitter in watts
(W) according to the transmitter ma-
nufacturer and d is the recommended
separation distance in meters (m).b
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey c,
should be less than the compliance
level in each frequency ranged.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m 80
MHz to
2.5 GHz
3 V/m
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is af
fected by absorption and reflection from structures, objects and people.
NOTE 3
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and landmobile radios, amateur radio, AM and FM radio bro
adcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electro
magnetic site survey should be considered. If the measured field strength in the
location in which the device is used exceeds the applicable RF compliance level
above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as
reorienting or relocating the device
NOTE 4
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
.
Recommended separation distances between portable and mobile RF communications
equipment and this device:
This device is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of this device can help prevent electromagne-
tic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and this device as recommended below, according to the
maximum output power of the communications equipment.