Select the kmh/kmh (1)
conversion for if you are using
the Trackmaster protocol.
If you are using a protocol other
than Trackmaster select "no
conversion" (0).
Select "Disable" (0) for Global
communication timeout.
Press "Y" to save the new
settings and restart the device.
Your device is now ready for use
with the external controller.
Note that the device is
automatically switched to
terminal mode when no manual
controller is present at the
device itself.
22/22
INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15