2. Press the enter key #4 until the display shows terminal 0
3. By using the "+" and "-" keys for the grade adjustment #3 you can change the "0" behind terminal to
"1" to activate the Terminal mode.
4. Press enter #4 and "prot id" will be shown on the display, for Lode protocol this should be set to "1".
Set the protocol as described in the "Ergometer setup" chapter.
The following protocols can be selected
Lode = Lode protocol
trAck = Trackmaster communication protocol
UUood = Woodway communication protocol
HPcoS = HP Cosmos communication protocol
gE = GE protocol
5. Press enter #4 and "baudrate" will be displayed to set the correct baudrate, select "0" for auto select.
The following baudrates can be set:
Baudrate = 0 (Autoselect dependent on protocol-id)
2400;
4800;
9600;
19200;
38400;
57600
6. Press enter #4 and "interface" will be displayed, you can choose the communication port which you
would like to select.
Interface = CIM interface choice. Select "RS232" or "USB" depending on the cable that you connected to
the treadmill.
0 = RS232
1 = RS232-AUX
2 = USB
7. Press enter #4; the next step "Neg Elev" is read only and is "1" when a negative elevation option has
been added to the treadmill, without a negative elevation option this value shows "0".
Neg Elev = Negetive elevation (read-only)
0 = not present
1 = present
Press the stop button #2 on the control unit or wait 10 seconds to leave the settings menu and let your
changes take effect.