8 - Treadmill or Bicycle without control unit

8 - Treadmill or Bicycle without control unit

8.1 - Device without control unit

8.1 - Device without control unit

In case the device does not have a control unit the correct protocol can be set from a computer by using

the Lode Service Manager tool that can be downloaded from the Lode website if you are a registered

user. http://lode.nl/en/product/getDownload/469

Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cable from the

computer to the RS232 port

Start the Lode Service Manager and login with the same account details that you use to login on the

website.

19/22

INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15

Summary of Contents for NIHON KOHDEN 1550K

Page 1: ...50K CPET Ergometer Service documentation Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t 31 50 571 28 11 f 31 50 571 67 46 ask lode nl www lode nl INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 2: ...3 3 Ergometer setup 4 Setting up the hardware 4 1 Connection overview 4 2 Connecting with the usb cable 4 3 Connecting with the RS232 cable 5 Treadmill or ergometer with 3 5 touch screen unit 6 Treadmill or ergometer with 7 touch screen control unit 7 Nummerical Control Unit Treadmill only 8 Treadmill or Bicycle without control unit 8 1 Device without control unit ...

Page 3: ...rm covers All Corival ergometers starting with serialnumber 9609xx 20150001 If your treadmill or ergometer is not covered by this interface form please contact Lode service service lode nl 3 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 4: ...em side is connected to the SIO 1 port If an internal blood pressure device will be used you only need a trigger cable under exceptional conditions In case the used ergometer is an Excalibur sport with blood pressure Or when on special request of a customer a external triggered blood pressure unit instead of standard blood pressure unit has been built in In these to cases you will need a trigger c...

Page 5: ...elect Register to store the selection Select Exam to store the selection and start a test The system is now ready for communication with the ergometer 3 3 Ergometer setup The following protocol needs to be set in the Lode ergometer to communicate with the external equipment Bicycle ergometer Ergoline P10 The next chapters describe the procedures to setup these protocols depending on the type of er...

Page 6: ...matically You can find the driver on the Lode website under the Services tab Locations For the Treadmill the communication ports are Located on the front side near the power switch For the bicycle ergometer the communication ports are located on the bottom of the bicycle Tilt the ergometer on it s side to be able to acces these ports Communication Module Use the single RS232 port to connect the co...

Page 7: ...onnect the cable to the Lode ergometer 3 Connect the cable to the external device 4 3 Connecting with the RS232 cable Connecting with the RS232 cable 1 Depending on the external device you may need to use a straight RS232 cable Lode part no 930911 or a cross linked cable The cable selection chapter specifies which cable is needed 2 Connect the cable to the Lode ergometer 3 Connect the cable to the...

Page 8: ...e the kind of device the serial number and the connected modules is displayed After initialisation the terminal mode is automatically displayed Swipe with your finger over the screen as indicated with the red arrow in the picture to go to the main menu 8 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 9: ...ver The menu Terminal Settings lets you set the communication settings for external communication The menu Service is only intended for service purposes to be executed by factory trainend personel Select Terminal Settings to change the communication protocol 9 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 10: ...Select the Communication protocol as desribed in Ergometer setup chapter Select Default for the baudrate 10 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 11: ... or USB depending in which cable you have connected to the Lode ergometer Select the button with the house to go back to the main menu 11 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 12: ...Select Terminal to set the ergometer in terminal mode Your device is now ready for external control 12 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 13: ... the Test window or main menu To access the menu from the Test window you need to swipe the graph area of the touch screen as indicated in the picture Note that for this instruction the graphic view for a Lode Corival is used The view for a treadmill might difer a bit but the procedure is the same 13 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 14: ...sing the red marked scroll button and select Terminal Select the communication protocol ad described in the Ergometer setup chapter Use the scroll button to view the complete list 14 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 15: ...Select Default for the baudrate Select RS232 or USB depending on the type of cable that you connected to the Lode ergometer 15 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 16: ...optional Your ergometer can be set to boot up in terminal mode Follow these steps to activate this function Go to the settings menu Select Default Start Menu Select Terminal Switch the ergometer off and on It will boot directly into terminal mode 16 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 17: ...7 Nummerical Control Unit Treadmill only 1 Press and hold key 1 and 2 as shown on Figure 1 for 10 seconds 17 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 18: ...uto select The following baudrates can be set Baudrate 0 Autoselect dependent on protocol id 2400 4800 9600 19200 38400 57600 6 Press enter 4 and interface will be displayed you can choose the communication port which you would like to select Interface CIM interface choice Select RS232 or USB depending on the cable that you connected to the treadmill 0 RS232 1 RS232 AUX 2 USB 7 Press enter 4 the n...

Page 19: ... be downloaded from the Lode website if you are a registered user http lode nl en product getDownload 469 Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cable from the computer to the RS232 port Start the Lode Service Manager and login with the same account details that you use to login on the website 19 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER V...

Page 20: ...ted 3 Select preset Lode Bus Service Protocol 4 Press OK 5 Press Break 6 Switch the device off wait 10 seconds 7 Switch the device on The screen as shown left should show up within 5 seconds after powering up 20 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 21: ...described in the Ergometer set up chapter Choose which port you want to use RS232 0 or USB 1 This setting refers to the type of cable that you have connected to the Lode ergometer Select default for the baudrate Note that Default should always be selected unless different specified by your dealer 21 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Page 22: ...Disable 0 for Global communication timeout Press Y to save the new settings and restart the device Your device is now ready for use with the external controller Note that the device is automatically switched to terminal mode when no manual controller is present at the device itself 22 22 INTERFACE SHEET NIHON KOHDEN 1550K CPET ERGOMETER VERSION 1 0 2016 01 15 ...

Reviews: