KELLY STAND
LOCO - 124 USER MANUAL
Page
9
Select Healthcare (UK) Ltd
Unit 2 Albert Road
Rushden
Northants
NN10 0BU Tel:
01933 411729 Fax: 01933 410213
Designed and Developed by: Proto Magic Innovations Ltd Page 9
KELLY STAND INSTRUCTIONS 07
7.
Trouble Shooting.
IF KELLY STAND FAILS TO LIFT OR LOWER CHECK THE FOLLOWING:
YES NO
a) Is the lift actuator's (18) power plug (35) plugged into the lift actuator power socket
(6) ? If NO plug in power plug (35) to connect lift actuator (18). If YES GO TO (b).
o o
b) Has emergency stop switch been activated? If YES pull out emergency
switch (2) to reset. If NO GO TO (c).
o o
c) Is battery (1) correctly located in the battery compartment(28)? If YES GO TO (d).
o o
d) Are any of the indicator lamps (31,32,33,34) on the battery compartment(28)
on? If YES GO TO (e). If NO then try a newly charged battery.
o o
e) Is green battery OK lamp (31) on the battery compartment(28) on? If YES GO TO (j).
o o
f) Is amber change battery lamp on (32) on the battery compartment(28) on? If YES
charge battery (1) or replace with fully charged battery if NO GO TO (d).
o o
g) Is KELLY STAND bleeping and red charge battery lamp on (33) on the battery
compartment(28)
on? If YES charge battery or replace with fully charged battery if NO GO TO (d).
o o
h) Is KELLY STAND bleeping and red overload lamp (34) on the battery compartment(28)
on? If YES press down button (19) on hand control (3) to reset then GO TO (l).
o o
j) Is hand control (3) plugged in securely? To hand control socket (15)
If YES GO TO (m).
o o
k) Has the maximum or minimum height been reached? If NO GO TO (m).
o o
l) Is sling caught on legs or other obstruction, or is client over safe working load
of 203kg (32 stone). If YES clear obstruction or a use higher capacity lifter
o o
m) If KELLY STAND is still not working,
CUSTOMER SERVICE ON (01933) 411729