KELLY STAND
LOCO - 124 USER MANUAL
Page
10
Select Healthcare (UK) Ltd
Unit 2 Albert Road
Rushden
Northants
NN10 0BU Tel:
01933 411729 Fax: 01933 410213
Designed and Developed by: Proto Magic Innovations Ltd Page 10
KELLY STAND INSTRUCTIONS 07
IF KELLY STANDLEG OPENING WILL NOT WORK CHECK THE FOLLOWING YES NO
a)Is the leg actuator power plug (12) plugged into the leg actuator power socket on battery
compartment (28) ? If NO plug in power plug (12) to connect leg actuator. GO TO (b).
o o
b) Make sure legs are not obstructed GO TO (c).
o o
c) Has emergency stop switch been activated? If YES pull out the emergency
stop button (2) to reset. If NO GO TO (d).
o o
d) Is battery (1) correctly located in the battery compartment(28)? If YES GO TO (e).
o o
e) Are any of the indicator lamps (31,32,33,34) on the battery compartment(28)
on? If YES GO TO (f). If NO then try a newly charged battery.
o o
(f) Is green battery OK lamp (31) on the battery compartment(28) on? If YES GO TO (j)
o o
(e) Is amber change battery lamp on (32) on the battery compartment(28) on? If YES
charge battery or replace with fully charged battery GO TO (f).
o o
f) Is KELLY STAND bleeping and red charge battery lamp on (33) on the battery
compartment(28) on? If YES charge battery or replace with fully charged battery
o o
GO TO (f).
g) Is KELLY STAND bleeping and red overload lamp (34) on the battery
compartment(28) on? If YES press down button (19) on hand control (3) to reset
then GO TO (j).
o o
h) Is hand control (3) plugged in securely? To hand control socket (15)
If YES GO TO (m).
o o
i) Are the legs fully open or closed? If NO GO TO (k).
o o
j) Are the legs caught on an obstruction, or is client over safe working load of 203kg
(32 stone). If YES clear obstruction or use a use higher capacity lifter if NO GO TO (k)
o o
k) If leg opening still does not work,
CUSTOMER SERVICE ON (01933) 411729