22
Instruction Manual
SECTION 6 - ADVANCED CONFIGURATION MENU
6. Press the POWER button. The pump will
begin pulsing. Adjust the stroke length knob to
the desired setting for calibration of point 2. Use
the UP or DOWN buttons to adjust the
stroke speed to the desired setting for calibration
of point 2. Press the POWER button when
adjustments are complete. The pump will stop
pulsing. The two stroke length settings must be
at least 10% apart. If not, a <10 error icon will
display.
7. Press the POWER button to begin calibration
of point 2. The pump will start pulsing. The
display will begin counting the strokes and
displaying the accumulated volume. The
volume displayed is based upon the pre-existing
settings. Press the POWER button when the
desired number of strokes has occurred.
8. Use the UP or DOWN buttons to edit the
highlighted volume as necessary. Press the
POWER button when complete.
9. Use the UP or DOWN buttons to select
whether to (Save) or (abandon) the
displayed calibration settings. Press the
POWER button to accept the selection and
then return to the main calibration menu.