SECTION 6 - ADVANCED CONFIGURATION MENU
6.6.1
Single-Point Calibration Procedure
During calibration, accuracy improves as stroke
count increases. As a result, a minimum of 100
strokes should be used to calibrate the pump.
The maximum number of strokes in the calibration
process is 9999. The stroke length cannot be
changed during calibration step 3 below. If the
stroke length is changed, a
SL
error icon will
display, the pump will stop, and the stroke count
and pump output will be reset to zero. Stroke speed
may only be adjusted in step 2. In general, the
MODE button may be used to step back in the
calibration procedure.
Procedure:
1. Upon selecting single-point calibration, the
icon will be displayed indicating that pump
is ready to begin calibration.
2. Press the POWER button. The pump will
begin pulsing. Adjust the stroke length knob
to the desired setting for calibration. Use the
UP or DOWN buttons to adjust the stroke
speed to the desired setting for calibration.
Press the POWER button when adjustments
are complete. The pump will stop pulsing.
3. Press the POWER button to begin calibration
of the pump. The pump will start pulsing.
The display will begin counting the strokes
and displaying the accumulated volume. The
volume displayed is based upon the pre-existing
settings. Press the POWER button when the
desired number of strokes has occurred.
4. Use the UP or DOWN buttons to edit the
highlighted volume as necessary. Press the
POWER button when complete.
5. Use the UP or DOWN buttons to select
whether to (Save) or (abandon) the
displayed calibration settings. Press the
POWER button to accept the selection and
then return to the main calibration menu.