background image

6

PROCEDURE FOR  ON-SITE VOLUMETRIC CALIBRATION E7 IN INTERNAL MODE 

A. 

Make certain speed control knob is turned fully clockwise. Turn off power to pump.

B. 

With pump primed and discharge tubing connected to the injection point as it would be in normal  

 

service, place foot valve and strainer assembly in a graduated container with a volume of 500 ml or  

 

more (so that the volume displaced by tubing and foot valve assembly is minimal in relation to    

 

volume of the solution in the container).

C. 

Turn pump on, and pump until air is exhausted from the suction line and pump head.

D. 

Turn pump off, note the solution level in the graduate. Refill graduate if necessary.

F.  

Turn off pump at the end of the calibration period, note volume pumped during the calibration    

 

period, and calculate volume of solution pumped in the time unit you choose (minute, hour, day,   

 

 etc.).

G. 

Adjust stroke frequency and/or stroke length knobs to your best estimate of required correction, and  

 

repeat  calibration measurements as a check.

 

PROCEDURE FOR  ON-SITE VOLUMETRIC CALIBRATION OF E7 IN EXTERNAL MODE 

A. 

Since pump output is governed by an external device, only the output per stroke may be calibrated.

B. 

With pump primed and discharge tubing connected to the injection point as it would be in normal  

 

service, turn off pump and place foot valve and strainer assembly in a graduated container with a  

 

volume of 500 ml or more.

 

C. 

Turn pump to internal mode with speed knob set at 100 and operate until air is exhausted from    

 

suction line and pump head.

 

D. 

Turn pump off and note solution level in graduated container. Refill graduate if necessary.

E. 

Turn pump on and count the number of strokes for exactly one minute. Then switch pump off.

F. 

Note volume pumped during the calibration period of one minute. Divide into this, the number of   

 

strokes to determine the volume of solution pumped per stroke.

 

 

Example:  50 ml/100 spm = 0.5 ml per stroke. 

 

 

 

Multiply this by your expected stroke rate per minute, per hour or per day and compare with  

 

 

desired output requirements

.

G. 

Adjust stroke length knob (lower knob) to your best estimate of required correction and repeat    

 

calibration procedure.

 

 

 

You may elect, for safety and convenience, to do the first calibration or operating test with water or other 
non-hazardous solution. If so, make certain the water or test fluid is removed from the Liquid End before 
pumping solutions that may react with the test fluid or be exothermic, such as sulfuric acid.

The final calibration adjustment should be made with pumping conditions identical to conditions of normal 
pumping service. This means that factors such as injection pressure, fluid viscosity, suction lift and others 
will automatically be accounted for in making the final adjustment of the pump.

   

Max. Pressure Rating

  

Max. Output

 Series    

psi

Bar

GPH

l/h

  E70 

300 

20.7 

1.3 

4.9

  E71 

150 

10.3 

2.5 

9.5

  E72 

100 

6.9 

4.0 

15.1

  E73 

60 

4.1 

8.0 

30

  E74 

25 

1.7 

20.0 

76

1718.D_306.indd   6

4/4/06   1:21:20 PM

Summary of Contents for E7 Series

Page 1: ...rdering replacement parts for your LMI Pump please include the complete Model Number and Serial Number of your unit Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 2...

Page 2: ...sories 5 Output Adjustment Initial Approximation 5 Procedure for On Site Volumetric Calibration E7 in Internal Mode 6 Procedure for On Site Volumetric Calibration E7 in External Mode 6 2 0 Troubleshoo...

Page 3: ...area that is convenient to both solution injection point and electrical supply LMI Series E7 metering pumps have corrosion resistant housings but should not be subjected to continuous high temperatur...

Page 4: ...is little or no resistance back pressure the velocity of the pumped fluid will be so great as to cause over pumping Be cause of this characteristic back pressure equal to approximately 25 psi 1 7 Bar...

Page 5: ...After installation is complete and an initial approximation setting has been made the pump should be calibrated and the stroke frequency and or stroke length settings adjusted F Nominal output and pre...

Page 6: ...lume of 500 ml or more C Turn pump to internal mode with speed knob set at 100 and operate until air is exhausted from suction line and pump head D Turn pump off and note solution level in graduated c...

Page 7: ...ump by turning the inner section of the knob counter clockwise Position the outer section of the knob so that the pointer aligns with zero on the nameplate Push down on the outer section of the knob a...

Page 8: ...ockwise rotation until motion stops Reset knob by grasping the knob and pulling it towards you Pry the yellow cap off of the knob Place the knob on a flat surface Using needle nose pliers squeeze the...

Page 9: ...iquifram setting is not correct restart pump then screw the Liquifram in or out Always stop pump electrically when checking Liquifram setting Repeat procedure if necessary After Liquifram is set prope...

Page 10: ...10 3 0 EXPLODED VIEW 28 29 29 30 27 26 25 24 23 31 33 41 34 32 42 37 1718 D_306 indd 10 4 4 06 1 21 49 PM...

Page 11: ...5839 Spacer 3 0 1 28 E7 10973 Shaft Seal 1 29 E70 E71 E72 E73 10598 Screw 10 24 S S 4 E74 10598 Screw 10 24 S S 8 30 E7 25452 Label 1 31 E7 31542 Nameplate 1 32 E70 E71 32465 Stroke Dial 1 E72 32466 S...

Page 12: ...on Roy All Rights Reserved Printed in USA Specifications subject to change without notice Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 215 293 0401 FAX 215 293 04...

Reviews: