background image

8

D.

 

Ruptured pumping diaphragm—

If rupture is severe and pump is injecting into pressure, solution  

 

leak will  be obvious through the 3/16” (5 mm) diameter hole at the bottom of the spacer directly   

 

behind the pump head. Replace pumping Liquifram™ (diaphragm). If rupture is a small pin hole,  

 

there may be oozing of solution through the 3/16” (5 mm) diameter hole described above. Replace 

 

ment of pumping Liquifram™ (diaphragm) will be necessary.

 

E.

 

Clogged Liquid End—

will cause low pump output. Purge Liquid End. Disassemble Liquid End.   

 

Clean individual parts with water and detergent or appropriate cleaning solution.

 

F.

 

Injection into excessive pressure—

If discharge pressure is within 25 psi (1.7 Bar)* of maximum  

 

pump rating, remove spring in injection check valve, if so supplied.

 

 

* 10 psi (0.7 Bar) for Series E74.

 

2.2    EXCESSIVE PUMP OUTPUT

A.

 

Syphoning

—If your Series E pump has a Liquid End other than a high viscosity LE, its injection  

 

check valve assembly has a chemically resistant anti-syphon spring. Disassemble the injection    

 

check valve and check to be sure this spring is in place and undamaged. Replace if necessary.   

 

 

Note that the anti-syphon spring must be removed if injection is into a pressure within 25 psi (1.7  

 

Bar)* of pump’s maximum pressure rating. Liquid Ends with a suffix “S” have, in addition, a dia-   

 

phragm type anti-syphon valve, which prevents syphoning and over pumping.

 

 

* 10 psi (0.7 Bar) for Series E74.

 

B.

 

Incorrect knob settings

—Check stroke length knob (lower knob) by rotating it counter-clockwise  

 

 to zero position. The Liquifram™ (pumping diaphragm) should stop reciprocating. If it does not,  

 

reset knob and continue counter-clockwise rotation until motion stops. Reset knob by grasping the  

 

knob and pulling it towards you. Pry the yellow cap off of the knob. Place the knob on a flat surface.  

 

Using needle nose pliers, squeeze the inner section together while lifting the outer section up (see  

 

Figure 5). Push the  inner section back onto the “D” shaped stroke shaft. With the pump running,  

 

zero the pump by turning the inner section of the knob counter-clockwise     . Position the outer    

 

section of the knob so that the pointer aligns with zero (0) on the nameplate. Push down on the   

 

outer section of the knob (a snap sound will indicate that parts are locked together). Replace the  

 

yellow  cap over the outer section of the knob, aligning the tabs on the cap with the slots on the   

 

knob.

 

2.3   ChANGING PUMPING LIQUIFRAM™ (DIAPhRAGM)

A. 

Make sure size code marking (0.9 S.I., 1.8 S.I. or 3.0 S.I.) on spacer and Liquifram™ (diaphragm)  

 

are the same. Liquifram™ and spacer size code must match in order for the pump to function. The  

 

E74 Liquifram™size code is 6.0 S.I., (spacer is 3.0 S.I., but there is a large black spacer adapter).

 

   

 

Always wear protective clothing, gloves and face shield when working on or near chemical metering  

 

pumps.

 

B. 

Depressurize discharge line following steps outlined on the supplement “Liquid Handling 

 

Assembly Instructions”. Lift foot valve from solution and let pump run pumping air for a few minutes.  

 

Then remove pump head.

C. 

Set stroke length knob (lower knob) to zero,  rotate it counter-clockwise      with the pump running, 
then stop the pump by turning stroke frequency knob (upper knob) to “off”.

D. 

Lift edge of Liquifram™ and rotate it counter-clockwise      to unscrew the Liquifram™.

E. 

Before installing new pumping Liquifram™, turn pump on and rotate stroke length control knob 
(lower knob) to the proper setting per the following table:

1718.D_306.indd   8

4/4/06   1:21:25 PM

Summary of Contents for E7 Series

Page 1: ...rdering replacement parts for your LMI Pump please include the complete Model Number and Serial Number of your unit Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 2...

Page 2: ...sories 5 Output Adjustment Initial Approximation 5 Procedure for On Site Volumetric Calibration E7 in Internal Mode 6 Procedure for On Site Volumetric Calibration E7 in External Mode 6 2 0 Troubleshoo...

Page 3: ...area that is convenient to both solution injection point and electrical supply LMI Series E7 metering pumps have corrosion resistant housings but should not be subjected to continuous high temperatur...

Page 4: ...is little or no resistance back pressure the velocity of the pumped fluid will be so great as to cause over pumping Be cause of this characteristic back pressure equal to approximately 25 psi 1 7 Bar...

Page 5: ...After installation is complete and an initial approximation setting has been made the pump should be calibrated and the stroke frequency and or stroke length settings adjusted F Nominal output and pre...

Page 6: ...lume of 500 ml or more C Turn pump to internal mode with speed knob set at 100 and operate until air is exhausted from suction line and pump head D Turn pump off and note solution level in graduated c...

Page 7: ...ump by turning the inner section of the knob counter clockwise Position the outer section of the knob so that the pointer aligns with zero on the nameplate Push down on the outer section of the knob a...

Page 8: ...ockwise rotation until motion stops Reset knob by grasping the knob and pulling it towards you Pry the yellow cap off of the knob Place the knob on a flat surface Using needle nose pliers squeeze the...

Page 9: ...iquifram setting is not correct restart pump then screw the Liquifram in or out Always stop pump electrically when checking Liquifram setting Repeat procedure if necessary After Liquifram is set prope...

Page 10: ...10 3 0 EXPLODED VIEW 28 29 29 30 27 26 25 24 23 31 33 41 34 32 42 37 1718 D_306 indd 10 4 4 06 1 21 49 PM...

Page 11: ...5839 Spacer 3 0 1 28 E7 10973 Shaft Seal 1 29 E70 E71 E72 E73 10598 Screw 10 24 S S 4 E74 10598 Screw 10 24 S S 8 30 E7 25452 Label 1 31 E7 31542 Nameplate 1 32 E70 E71 32465 Stroke Dial 1 E72 32466 S...

Page 12: ...on Roy All Rights Reserved Printed in USA Specifications subject to change without notice Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 215 293 0401 FAX 215 293 04...

Reviews: