background image

7

2.0  TROUBLEShOOTING LIQUID END

2.1  LOW PUMP OUTPUT

Low pump output can be caused by many things. Some of the more common ones are: 
 

•  Very low stroke setting, i.e. red zone setting of knob

 

 

•  Trapped air in pump head (trapped air in discharge tubing has no effect)

 

 

•  Air leak through valve seal rings

 

 

•  Ruptured pumping Liquifram™ (diaphragm)

 

 

•  Clogged Liquid End, or injection point connection

 

 

•  Injection into pressure within 25 psi (1.7 Bar)* of pump’s maximum pressure rating with anti-   

 

  syphon spring in place (if so supplied)

 

 

•  Injection into pressure in excess of pump rating-see chart on page 6.

 

 

* 10 psi (0.7 Bar) for  Series E74.    

 

A.

  

Very low stroke setting

 - Check position of stroke length knob (lower knob) by rotating it counter-

clockwise     until Liquifram™ (diaphragm) stops moving with the pump operating. The Liquifram™ 

should not stop reciprocating (moving or clicking) until the knob points to zero.  If it stops before 

zero, reset knob by grasping the knob and pulling it towards you. Pry the yellow cap off of the knob. 

Place the knob on a flat surface. Using needle nose pliers, squeeze the inner section together while 

lifting the outer section up (see Figure 5, below).

 

 

   

 

Push the  inner section back onto the “D” shaped stroke shaft. With the pump running, zero the 
pump by turning the inner section of the knob counter clockwise. Position the outer section of the 
knob so that the pointer aligns with zero on the nameplate. Push down on the outer section of the 
knob (a snap sound will indicate that parts are locked together). Replace the yellow cap over the 
outer section of the knob, aligning the tabs on the cap with the slots on the knob.

B.

 

Trapped air in pump head - 

May be caused by leaks in the suction line, where the suction line 

joins the suction fitting, or at the seal ring between suction fitting and pump head. It may also be 
caused by air or gases coming out of solution. Trapped air or bubbles in the discharge line have no 
effect on the pump’s operation. They may be ignored.

 

 

•  To remove trapped air from the pump head, operate the pump with stroke frequency knob and 
stroke length knob set at 100.

 

•  It may be neccessary to pull on both knobs of the 4FV, or rotate pressure relief knob ¼ turn, if 
so equipped. Follow “Priming” instructions in the Liquid Handling Assembly Sheet inserted in this 
instruction book and operate the pump for a few minutes to purge the head and valves of air or gas.

C.

 

Air leak through valve seal rings-

 usually caused by worn or damaged seal rings or loose fittings. 

Tighten fittings by hand until they are very snug. If there is no improvement, replace both seal rings 
in pump head. See enclosed Liquid Handling Assembly Sheet.

 

Figure 5

 

 

 

 

 

 

1718.D_306.indd   7

4/4/06   1:21:24 PM

Summary of Contents for E7 Series

Page 1: ...rdering replacement parts for your LMI Pump please include the complete Model Number and Serial Number of your unit Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 2...

Page 2: ...sories 5 Output Adjustment Initial Approximation 5 Procedure for On Site Volumetric Calibration E7 in Internal Mode 6 Procedure for On Site Volumetric Calibration E7 in External Mode 6 2 0 Troubleshoo...

Page 3: ...area that is convenient to both solution injection point and electrical supply LMI Series E7 metering pumps have corrosion resistant housings but should not be subjected to continuous high temperatur...

Page 4: ...is little or no resistance back pressure the velocity of the pumped fluid will be so great as to cause over pumping Be cause of this characteristic back pressure equal to approximately 25 psi 1 7 Bar...

Page 5: ...After installation is complete and an initial approximation setting has been made the pump should be calibrated and the stroke frequency and or stroke length settings adjusted F Nominal output and pre...

Page 6: ...lume of 500 ml or more C Turn pump to internal mode with speed knob set at 100 and operate until air is exhausted from suction line and pump head D Turn pump off and note solution level in graduated c...

Page 7: ...ump by turning the inner section of the knob counter clockwise Position the outer section of the knob so that the pointer aligns with zero on the nameplate Push down on the outer section of the knob a...

Page 8: ...ockwise rotation until motion stops Reset knob by grasping the knob and pulling it towards you Pry the yellow cap off of the knob Place the knob on a flat surface Using needle nose pliers squeeze the...

Page 9: ...iquifram setting is not correct restart pump then screw the Liquifram in or out Always stop pump electrically when checking Liquifram setting Repeat procedure if necessary After Liquifram is set prope...

Page 10: ...10 3 0 EXPLODED VIEW 28 29 29 30 27 26 25 24 23 31 33 41 34 32 42 37 1718 D_306 indd 10 4 4 06 1 21 49 PM...

Page 11: ...5839 Spacer 3 0 1 28 E7 10973 Shaft Seal 1 29 E70 E71 E72 E73 10598 Screw 10 24 S S 4 E74 10598 Screw 10 24 S S 8 30 E7 25452 Label 1 31 E7 31542 Nameplate 1 32 E70 E71 32465 Stroke Dial 1 E72 32466 S...

Page 12: ...on Roy All Rights Reserved Printed in USA Specifications subject to change without notice Replaces same of Rev C 10 97 1718 D 3 06 201 Ivyland Road Ivyland PA 18974 USA TEL 215 293 0401 FAX 215 293 04...

Reviews: