10
ENG
3. Press the M1 (or M2) Button to
see the
fi
gures stored in the memory.
At the first depression of button
M1 (or M2) mean value of 3 last
measurements, kept in memory M1
(or M2), will appear on the screen
(Fig. 6). At repeated depression
of button M1 (or M2) indicator of
selected memory M1 (or M2) and
number of memory cell will appear
on the screen, and in 1 second its content is displayed (Fig. 7). When content
of memory cell is displayed, date and time of measurement are displayed
alternately in the display lower line. Each depression of button M1 (or M2)
causes shifting to the next memory cell.
If there are no saved results in memory, at
fi
rst depression of button M1 (or
M2) current date and time are displayed on display Device.
MEMORY CLEARING
To delete all values stored in the memory, press the M1 (or M2) Button and
hold it down for more than 5 seconds. Symbols “Clr” will appear on the
screen and all values will be deleted from the memory.
INFORMATION ABOUT ERRORS
Indication
Likely cause
Methods of correction
The arm cuff is applied
incorrectly or the air tube plus
is inserted not tightly enough.
Measurements cannot be made
due to hand movement or
talking during measurements.
Be sure that the arm cuff is
applied correctly and the plug
is inserted tightly. Repeat the
whole measurement procedure.
Repeat the measurement fol-
lowing strictly the recommen-
dations of this Manual.
Batteries are discharged.
Replace the batteries for
new ones
CARE, STORAGE, REPAIR AND DISPOSAL
1. Keep this Device from exposure to higher humidity, direct sunlight,
shocks, vibration. THIS DEVICE IS NOT WATERTIGHT.
2. Do not keep and use this Device near heating installations and open
fi
re.
3. If the Device was stored at a temperature below the freezing point, keep
it at least for 1 hour in some warm place before use.