Originally
Adopted:
07/13/2006
Revision #:
1.1
Revision Effective Date: 09/08/2006
In addition, the following suggested corrective action is provided: 1) The operator should consider
removing any contaminants from the area (to include the subject, if necessary). 2) The operator
may also consider checking the instrument’s connection to the standard delivery system on the
right-hand side of the instrument. 3) Use a new mouthpiece and restart the test. 4) If the
instrument gives an additional exception message, consider placing the instrument out of service
by placing an “Intoxilyzer Out Of Service” form on the instrument and reporting the problem to the
OSP Regional Dispatch Center at 1-800-292-6517. 5) Leave the instrument power on and
consider locating an alternative testing location. 6) Any observations made by the Operator during
the testing sequence may be recorded in the comments section.
Diagnostic Failure
This exception occurs when the instrument fails the Diagnostic routine. The instrument displays
“Diagnostic Fail”
and prints
“*Invalid Test – Diagnostic Fail”
. The test is not complete and can
be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should check for
any obvious reason for the condition (e.g., did the subject blow at the wrong time?) 2) Use a new
mouthpiece and restart the test. 3) If the instrument gives an additional exception message,
consider placing the instrument out of service by placing an “Intoxilyzer Out Of Service” form on
the instrument and reporting the problem to the OSP Regional Dispatch Center at 1-800-292-
6517. 4) Leave the instrument power on and consider locating an alternative testing location. 5)
Any observations made by the Operator during the testing sequence may be recorded in the
comments section.
Deficient Sample
This exception occurs when the subject did not blow long enough with sufficient force to properly
provide a sample during the breath sample period. The instrument will display
“Deficient
Sample”
and print
“*Invalid Test – Deficient Sample”
. The test is not complete and can be
restarted unless the operator determines that the subject is physically incapable of providing a
proper sample.
In addition, the following suggested corrective action is provided: 1) The operator should consider
re-instructing the subject on how to provide a proper breath sample. 2) If the operator is continuing
with the testing process, use a new mouthpiece and restart the test. 3) Continue the test until the
subject properly submits to or refuses the test. 4) Any observations made by the Operator during
the testing sequence may be recorded in the comments section.
ALL PRINTED COPIES ARE UNCONTROLLED AND MAY NOT BE COMPLETE
Page 37 of 54