To change the displayed vacuum value from Bar to MPa simply touch the desired radio button.
Motor Tab
T96 System Controllers connected to stages which have motorised movement, such as the MDS600, TST350 or the GS350 require a number of motor
drive option boards, one for each motor.
Each motor tab will show the detail of the board and how it is configured and which stage is connected.
Extra information about the end stops and encoders are also detailed.
DSC Tab
DSC stages requires an option board with further signal conditioning circuitry for the highly sensitive DSC signal. This tab does not have any settings
a user may change and only details the stage it is connected to and also details of the option boards hardware.
Sample Calibration
Depending on the specific properties of your sample (for example; total mass, thermal conductivity, surface area and flatness), it is possible that
thermal losses between the heating block and the sample or a temperature gradient through the sample may cause a difference between the
reported temperature of the stage and the actual temperature of your sample. To account for this, the Sample Calibration function has been
included with the Linkpad or LINK software to allow you to apply an offset to the displayed temperature, if required.
To use sample calibration tap on the settings bar and select the 'Stage' tab and the following screen will appear:
Tap the 'Values' button and a further screen will appear:
To reset any stored values to the default settings tap the 'Clear' button which will cause a confirmation prompt to appear:
tap 'OK' and the stored values will be cleared, taping on cancel will leave the values as they are currently set.