32
Switch on the controller and reconnect to
Linksys32.
Click:
File > Connect
The Status window may display the message ‘The
gap motor might need to be referenced’
DO NOT
click on the reference ‘ref’ button or the Lid will
move up (2500mm) and the stage will have to be
manually reset to zero point again.
Using the PC mouse, right click onto the CSS450
toolbar.
Click and select:
Reference
Click ‘
OK
’ to continue.
There are three sensors pins located on top of the
Lid. They are labelled R (reference), Z (zero) and
L (lid).
Place the Sensor Adjustment Tool on the zero
sensor pin (Z).
Turn the Sensor Adjustment Tool so that the
‘Zero’ light in the CSS450 control panel just turns
green.
One turn of the screw represents a gap of 500
mm
.
So to get an accuracy of 1
mm
, it needs to be
within 0.72 degrees rotation.
When Zero is set, click on the ‘
Zero
’ button. The
Lid will now move up 2500mm to the Reference
point.