![Linkam Scientific Instruments CCR1000 User Manual Download Page 7](http://html.mh-extra.com/html/linkam-scientific-instruments/ccr1000/ccr1000_user-manual_1923492007.webp)
7
Mounting Stage to Microscope
with Dovetail Substage
The CCR1000 is designed to work with an up-
right microscope using reflected light only.
The following description is for mounting the
stage on to microscopes which have a circular
dovetail substage assembly.
Attach the curved stage clamps (part no. 9548) to
the base of the stage using the supplied hex
screws and the outer most holes in the base plate.
Adjust the two positioning screws (1) and the
thumb screw (2) so that approximately 5mm of
thread is exposed on the inside edge of the clamp.
This will roughly position the stage in the centre
of the dovetail (3).
Place the stage onto the dovetail then tighten the
thumbscrew (2) to fix the stage in place.
For other types of microscope sub-stages, refer to
the diagram included with the stage adaptor.
Curved Clamps Set 9548
1
3
2