Important Notice
Please check that your Linkam equipment has not been damaged during transport. If there is any evidence
of external damage DO NOT SWITCH ON ANY ELECTRICAL ITEMS.
Contact LINKAM SCIENTIFIC or their appointed distributor immediately. Your warranty may be im-
paired if Linkam is not informed of any transport damage within 7 working days of delivery.
NO attempt should be made to repair or modify the equipment in any way, as there are
no user replaceable
parts
.
No attempt should be made to open the case except by qualified personnel as hazardous voltages are pre-
sent.
Please contact Linkam for custom modifications for specific applications.
In order to use this equipment successfully, please take time to read this manual all the way through before
starting to work.
Warranty
This equipment has a warranty against defects in material and workmanship for a period of 12 months.
Linkam will either repair or replace products that prove to be defective. For warranty service or repair,
this product must be returned to Linkam or a designated service facility.
The warranty shall not apply to defects resulting from interfacing, unauthorized modification or misuse,
operation outside of the environmental specifications for the product, or improper site preparation or
maintenance.
Technical Support
Any technical questions or queries should be addressed to the Technical Support Department at the ad-
dress shown on the back of this manual.
Equipment Maintenance
The programmer does not require any regular maintenance. If for any reason it is necessary to check the
electronic calibration then a set of standard resistances can be supplied, which simply plug into the pro-
grammer in place of the stage, and indicate known temperature values. The standards are traceable to
NAMAS (National Accreditation of Measuring and Sampling)
Before cleaning the case or front panel of the programmer, remove the mains lead from the wall outlet.
Use a small quantity of isopropyl alcohol with a soft cloth and gently wipe the surface.
Pressure Warning
The CCR1000 has been tested to 5bar pressure using 22mm quartz windows of 1mm thickness. Windows
must always be inspected with magnifying lens to check there are no microscopic cracks or scratches
before using the stage under pressure. Never stand within 1m of CCR1000 when using high pressure
between 2-5bar.
Linkam cannot guarantee pressure with IR transparent windows due to the inconsistent mechanical
strength of the windows.
Contact Linkam if there is any concern with pressurizing this stage.