Load A sample
Remove Sample Holder
Before removing the sample holder please ensure
that the heater assembly has cooled to room tem-
perature. When using the Loading Tool, care
must be taken as the heater assembly is very
fragile.
Remove lid by unscrewing anti clockwise.
Remove the 22x1mm quartz window.
Use the Sample Holder Removal Tool (1) as
shown in the picture.
Carefully lower the Sample Holder Removal Tool
over the Sample Holder (2).
The next step requires great care and minimal
force. Once the tool is axially aligned with the
sample holder it will need to be pressed over the
outer diameter of the Sample Holder. Rotate the
tool slightly to make this easier. The tool will
come to a stop after approximately 1.5mm. No
further force is required otherwise the user risks
damaging the heater assembly. Gently pinch the
tool with a thumb and fore finger. This will allow
the tool to grip the outer diameter of the sample
holder.
Lift the tool vertically away from the CCR1000.
If the sample holder is lifted out at an angle it
could break.
1
2
Lift Vertically