D9U001L20-0101_04
34
Trendelenburg Position
The Trendelenburg Position helps create anti-shock conditions for the patient.
Settings after pressing position button 21:
■
All mattress platform elements are retracted to their lowest position.
■
The mattress platform is inclined by up to 16° (head lower than feet).
The bed moves into the Trendelenburg position within 30 seconds from any mattress platform position. You cannot disable this
function.
Examination Position
The examination position is suitable for:
►
Examining the patient
►
Performing a procedure on the patient
►
Moving the patient to another bed
Settings after pressing position button 19:
■
The backrest and thighrest are moved to the horizontal position.
■
The mattress platform is moved to its highest position.
You can disable this function. If individual functions (for example, thighrest adjustment) are disabled, they are not performed.