
D9U001L20-0101_04
20
6.4 Electromagnetic compatibility
Bed is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical system for mag
-
netic resonance imaging, where the intensity of EM disturbances is high.
Bed has defined no essential performance.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead to
improper operation. If such use is needed, this device and the other equipment should be under surveillance
to verify proper operation.
List of used cables:
1.
Mains cable, maximum length 6 m
2.
ACP Attendant Control Panel, maximum length 3m
3.
Handset, maximum length 3m
WARNING!
Use of the accessories, converters and other cables, than specified and provided by manufacturer of this bed
could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this bed and lead
to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should
not be used closer than 30 cm (12 inches) from any part of this bed Latera, including cables specified by manu
-
facturer. Otherwise this could lead to deterioration of functionality of this bed.
WARNING!
Do not overload the bed (SWL), respect the duty cycle (INT.) and consider chapter 18 Maintenance in order to
maintain the basic safety with regard to electromagnetic disturbances for the expected service life.
6.4.1 Manufacturer instructions - electromagnetic emissions
Emission Test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies