D9U001GT0-0101_09
45
12.2.2 Activating GO Button
The GO button 17 activates the keypads on all control elements except the foot control. A GO button is included on a number of
different control elements. The function of the GO button is identical on all control elements.
After pressing GO button 17, the keypad will remain active for 3 minutes. It is possible to control every function on the bed except
the locked functions.
Pressing a function button will keep the keypad active for another 3 minutes.
It is necessary to activate the keypad again if the 3 minute period without pressing any function is passed.
12.2.3 Function Buttons
Function buttons 3, 4, 5, 6, 9 and 10 allow the setting of different positions, for example the height and tilt of the mattress plat-form,
adjust individual mattress platform elements, etc.
To set a position:
►
Activate the keypad by pressing the GO button.
►
Press and hold the respective button until required position is reached.
12.2.4 Lock Buttons
Lock button 2 allows disabling individual functions on the Attendant Control Panel.
To disable functions:
►
Activate the keypad by pressing the GO button 17.
►
Press respective lock button 2.
The respective LED flashes 1 to indicate the lock.
NOTE:
The individual functions are locked in the central control panel, the satellite control, the handset, and the siderail control.
The foot controllers are locked separately.
12.2.5 Mains Power LED
Status
Meaning
lit LED
connected to the mains
unlit LED
disconnected from the mains
flashing LED
system error
12.2.6 Position Buttons
WARNING!
Risk of injury due to moving parts!
►
Ensure that no body parts are trapped between moving parts of bed and mattress platform.
►
Ensure that no persons or body parts are close to bed or accessories (e.g. infusion stand, lifting pole) when
mattress platform is moving
CAUTION!
Damage to property due to moving parts!
►
Ensure that no objects (e.g. cables) are trapped between moving parts of bed and mattress platform.
►
Ensure that no objects are close to bed or accessories (e.g. infusion stand, lifting pole) when mattress platform is
moving.
The therapeutic and safety-related positions are pre-programmed. When a position is set, several parts of the bed and mattress
platform will move simultaneously.
Programmed positions:
►
Cardiac chair position 13
►
Trendelenburg position 12
►
CPR (resuscitation) position 11
To set the programmed positions:
►
Activate the keypad by pressing the GO button.
►
Press and hold corresponding function button until the desired position is reached.