D9U001GT0-0101_09
28
Table 1 - IMMUNITY to RF wireless communications equipment
Test frequency (MHz)
Band (MHz)
Service
Modulation
Immunity Test Level
V/m
385
380 - 390
TETRA 400
Pulse modulation 18 Hz
27
450
430 - 470
GMRS 460, FRS 460
FM ± 5 kHz deviation 1
kHz sine
28
710
745
780
704 - 787
LTE band 13, 17
Pulse modulation 217
Hz
9
810
870
930
800 - 960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850,
LTE band 5
Pulse modulation 18 Hz
28
1 720
1 845
1 970
1 700 - 1 990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE band 1, 3,
4, 25; UMTS
Pulse modulation 217
Hz
28
2 450
2 400 - 2 570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450,
LTE band 7
Pulse modulation 217
Hz
28
5 240
5 500
5 785
5 100 - 5 800
WLAN 802.11 a/n
Pulse modulation 217
Hz
9
NOTE
There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE
There are no known other measures for keeping the basic safety based on EMC phenomena.
NOTE
Beds equipped with communication module meet standard for IEEE 802.11 b/g/n (2400,0 MHz – 2483,5 MHz, modulation
DSSS (IEEE 802.11 b ), OFDM (IEEE 802.11 g/n) 20MHz bandwidth, EIRP = 0,34 W).
7 Use and Storage Condition
The bed may not be used and stored in indoor environments:
■
Where there is a risk of explosion.
■
Containing inflammable anaesthetics.
The bed is designed for use in rooms for medical purposes. Electrical installations must therefore meet local standards laying down
the necessary conditions for electrical installations.
► Disconnect the bed from the mains in exceptional cases (i.e. earthquake, lightning).