2 Safety Instructions
WARNING!
Eleganza 1 bed should be left in its lowest position when the patient is unattended in order to reduce risk of
injury due to falls!
WARNING!
Siderails of Eleganza 1 should be located in the „up“ position to reduce the risk of the patient accidentally
slipping or rolling off the mattress!
WARNING!
Incompatible siderails and mattresses can cause an entrapment hazard!
WARNING!
Inappropriate handling of the power supply cord, e. g. by kinking, shearing or other mechanical damages is
hazardous!
WARNING!
When routing cables from other equipment in the Eleganza 1 bed avoid squeezing those between parts of the
Eleganza 1 bed!
WARNING!
Eleganza 1 bed should not be used with bed hoists and bed lifts!
WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
WARNING!
During specific investigations or treatments the significant risks of reciprocal interference posed by medical
electrical equipment may occur.
WARNING!
No modification of this equipment is allowed.
WARNING!
Do not modify this equipment without authorization of the manufacturer.
WARNING!
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe
use of the equipment.
WARNING!
An additional multiple socket-outlet or extension cord shall not be connected to the medical electrical system.