D9U001GT0-0101_09
11
ACCUMULATOR INDICATOR
MEANING
lit
accumulator disconnected or faulty
flashing: 1,6s lit / 0,2s not lit
accumulator deeply discharged
flashing: 0,1s lit / 0,1s not lit
accumulator discharged
flashing: 0,2s lit / 1,6s not lit
accumulator is charging
not lit
accumulator charged
1.9 Visual signalisation
1.9.1 Mains Power LED (Attendant Control Panel)
1.9.2 Accumulator Indicator (Attendant Control Panel, MiniACP)
1.9.3 Lock LED (Attendant Control Panel, MiniACP)
MAINS POWER LED
MEANING
lit
connected to the mains
flashing: 0,6s lit / 0,6s not lit
keyboard error (flashing inverted to Lock LED)
error (first fault)
flashing: 0,1s lit / 0,1s not lit
service mode
not lit
disconnected from the mains power
transformer switching error
VISUAL
SIGNALISATION
LOCK LED
lit
flashing:
0,6s lit / 0,6s not lit
not lit
Thighrest Lock LED
locked
lock
error
keyboard
error
motion
blocked
unlocked
Backrest Lock LED
locked
lock
error
keyboard
error
motion
blocked
unlocked
Bed Height, Trendelenburg and An-
titrendelenburg Tilt Lock LED
locked
lock
error
keyboard
error
motion
blocked
unlocked
Foot Control Lock
LED
locked
lock
error
keyboard
error
motion
blocked
unlocked