1.7 Definitions
Basic Bed Configuration
the pricelist model configuration, not including a mattress
Bed Weight
The value depends on the product configuration, accessories or customer
adjustments.
Clearance of Undercarriage
the height from the floor to the lowest point of the undercarriage between
the castors, for the manipulation of accessories under a braked bed in the
standard position
Ergoframe
Ergoframe is the kinematic system of Mattress support platform Adjustment
whose effect is the elimination of pressure on the patient´s abdomen and pel-
vic area and frictional forces on the patient‘s back and legs.
Maximum Patient Weight
Maximum Patient Weight depends on the application environment according
to IEC 60601-2-52. For application environment 1 (intensive/critical care) and
2 (acute care) reduce Safe working Load by 65 kg. For application environ
-
ment 3 (long-term care) and 5 (ambulatory care) reduce Safe working Load
by 35 kg.
Safe Working Load
the highest allowable load on the bed (patient, mattress, accessories and the
load supported by those accessories)
Siderail Height
the height of the upper crossbar or the edges of the siderails (not the highest
point of the siderail controls) from the patient surface
Standard Bed Position
■
The lowest height of the mattress support platform with regard to the floor
■
The mattress support platform, including the individual parts, has to be in a
horizontal (level - 0°) position.
■
The siderails are always locked in the upper position.
■
The basic position of the integrated extension.
Adult
Patient having a physical size equal to or more than 146 cm, a mass equal
to or more than 40 kg and a body mass index (BMI) equal to or more than 17
(according to IEC 60601-2-52).
CE
European Conformity
CPR
Cardiopulmonary Resuscitation
EAC
Eurasian Conformity
dB
Sound Intensity Unit
EMC
Electromagnetic Compatibility
FET
Field-effect transistor
HF
High Frequency
ICU
Intensive Care Unit
INT.
Duty Cycle
IV
Intravenous
LED
Light Emitting Diodes
ME
Medical Electrical (Equipment)
OFF
Deactivated
ON
Activated
REF
Reference number (product type depending on configuration)
SCU
System Control Unit
SN
Serial number
SWL
Safe Working Load
UDI
Unique Device Identification (for medical devices)
WEEE
Waste Electrical and Electronic Equipment