7 Use and storage Conditions 

The bed may not be used and stored in indoor environments:

■   

Where there is a risk of explosion.

■ 

Containing inflammable anaesthetics.

The bed is designed for use in rooms for medical purposes. Electrical installations must therefore meet local standards laying down 

the necessary conditions for electrical installations.

►         Disconnect the bed from the mains in exceptional cases (i.e. earthquake, lightning).

8 Scope of Delivery and Bed Variants

8.1 Scope of Delivery

Delivery:

 

The bed is delivered with dismounted head board and foot board, head bar and undercarriage. It is necessary to mount 

those parts to bed (according the chapter Assembly of Disassembled Bed). Upon receipt check that the shipment is complete as 

specified on the delivery note. Notify the carrier and supplier of any deficiencies or damages immediately as well as in writing or 

make a note on the delivery note.

8.2 Bed Variants

Features – Praktika 1 1F (model number is located on the serial label)

s = standard

o = optional

  

Mattress support platform

   

   Mattress support platform consisting of steel parts (s)

   

   Mattress support platform consisting of removable plastic parts (o)

  

Siderails

   

   Without siderails (s)

   

   Split plastic siderails (o)

   

   Single collapsible siderails (o)

  

Head board and Foot board

   

   Removeable head board and foot board with coloured inserts (s)

  

Castors

   

   125 mm Tente Motion with individual brake (s)

   

   125 mm Tente Motion with single axle brake (o)

  

Control Elements

   

   Controlling levers for back / foot rest setttings (s)

  

Other

   

   One pair of universal accessory holder (o)

   

   Segufix holders (o)

   

   Protector® (o)

   

   Extender® (o)

   

   Lifting pole holder (s) (head end)

   

   Accessory holder (o) (foot end)

  

Powder Coating Colours

   

   Powder coated steel parts – RAL 9002 (white) (s)

   

   Powder coated steel parts – RAL 9006 (grey) (o)

 

Summary of Contents for D9U001F10-0101

Page 1: ...Instructions for Use and Technical Description Praktika 1 Positionable hospital bed D9U001F10 0101 Version 06 Publication Date 2020 02...

Page 2: ...ET 2020 All trademarks and brands are the property of the appropriate owners The manufacturer reserves the right to changes in the contents of this manual that relate to the product s technical regula...

Page 3: ...lation with Levers 28 11 3Bed Positioning 28 11 3 1Backrest Positioning 28 11 3 2Thighrest Positioning 29 11 3 3Mechanical positioning of calfrest 30 11 4Quick Backrest Lowering for the Crisis Situati...

Page 4: ...arns about the risk of material damage WARNING warns about the risk of physical injury DANGER warns about the risk of fatal injury 1 1 2Structure of Warning Notices SIGNAL WORDS Type and source of dan...

Page 5: ...D9U001F10 0101_06 5 1 4Symbols on the Package FRAGILE HANDLE WITH CARE THIS WAY UP KEEP DRY PROTECT FROM HUMIDITY PAPER RECYCLING SYMBOL DO NOT USE HAND TRUCK HERE DO NOT STACK DURING STORAGE...

Page 6: ...ER DO NOT PLACE LIFTING POLE IN THIS PLACE SAFE WORKING LOAD SWL MAXIMUM WEIGHT OF PATIENT WEIGHT OF BED ONLY SUITABLE FOR INDOOR USE PROTECTION AGAINST ACCIDENTS DUE TO ELECTRICAL CURRENT TYPE B APPL...

Page 7: ...U001F10 0101_06 7 CE MARK OF CONFORMITY WITH EU REGULATION DESIGNATION OF HOSPITAL BED FOR ADULTS RECYCLING SYMBOL DO NOT POLLUTE THE ENVIRONMENT MEDICAL DEVICE compatible with Medical Device Regulati...

Page 8: ...7 W42 Praktika 1 REF SN ORDER 20190102326 20000290 2020 02 04 01 8592654208113 11 200204 21 20190102326 N A CUC 1 5 6 7 8 9 4 2 3 1 Address of Manufacturer 2 Manufacturing Date Year Month Day 3 DI Dev...

Page 9: ...he load supported by those accessories Siderail Height the height of the upper crossbar or the edges of the siderails not the highest point of the siderail controls from the patient surface Standard B...

Page 10: ...ts and bed lifts WARNING To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth WARNING During specific investigations or treatments the signific...

Page 11: ...ent authority of the Member State in which the user and or patient is established WARNING This medical device is not intended for oxygen enriched environment WARNING This medical device is not intende...

Page 12: ...rs are braked when the bed is not being moved regardless of whether the bed is occupied or empty Ensure that siderails are operated by healthcare personnel only Never use the bed in areas where there...

Page 13: ...ses doctors technical personnel transport personnel cleaning personnel 3 2Contraindications The medical device is not intended for the pediatric patients use Certain positions are not suitable for spe...

Page 14: ...Head split plastic siderail in highest positions 4 Head Board 5 Head siderail release mechanism 6 Protective bumper 7 Individual castor brake 8 Castor 125mm 9 Center siderail release mechanism 10 Acce...

Page 15: ...ble siderail in highest positions 3 Head Board 4 Protective bumper 5 Individual castor brake 6 Castor 125mm 7 Accessory rail 8 Siderail release mechanism 9 Controlling lever for backrest positioning 1...

Page 16: ...bed split plastic siderails down up single collapsible siderails down up 28 32 cm 32 32 cm Mattress support platform height 125 mm castors fixed height of 55 cm Maximum angle of the backrest 70 Maximu...

Page 17: ...Notify the carrier and supplier of any deficiencies or damages immediately as well as in writing or make a note on the delivery note 8 2Bed Variants Features Praktika 1 1F model number is located on t...

Page 18: ...and maintenance WARNING Material damage due to incorrect assembly Ensure that putting into service is performed exclusively by manufacturer s customer service or trained hospital personnel Set up the...

Page 19: ...and there are no accessories atta ched to the head board and foot board If a patient is lying in a bed with the head board and or foot board removed supervise the bed at all times WARNING Material dam...

Page 20: ...rs and its braking Check if it is possible to remove head board and foot board Check the functionality of siderails 9 4Transport Observe the following for safe transport Ensure the castors are unlocke...

Page 21: ...ndercarriage with castors from the bottom side of mattress support platform When removing the binding ribbons always hold the part which is being removed and place it on the clean and non abrasive sur...

Page 22: ...cing to the floor Torque 8 2Nm Using 4 carriage bolts M8x65 fix undercarriage to the head end of the bed Nut and distance washer are facing to the floor Torque 8 2Nm Cut off two straps on the sides wh...

Page 23: ...on the second side Turn the castors into the direction of tilting Brake the castors Ensure that second and third person lifts up the head siderails during tilting the bed on the castors to avoid dama...

Page 24: ...emove the nuts distance washers and screws Keep the screws for future use Remove the fixing thorns Insert the head bar inside the frame so the holes for screws bears on Fix the head bar to the bed usi...

Page 25: ...erails to the frame by removing the screws Non following this instructions may lead to damaging the siderails or mechanism of siderails Insert the head bar into the frame so the holes for screws bears...

Page 26: ...inding ribbons which fix head board and foot board to the mattress support platform 4 for head board and 4 for foot board Unpack the head board and foot board from the foil Insert the head board and f...

Page 27: ...lever thighrest positioning 2 Control lever backrest positioning Fig Control levers 11 1Positions of Levers WARNING Risk of damaging the bed during transport Always set both levers into transport posi...

Page 28: ...jury while positioning the bed Ensure no objects are trapped between moving parts of the bed and mattress support platform Ensure no objects are trapped between moving parts of the mattress support pl...

Page 29: ...backrest Fig Lowering the backrest 11 3 2Thighrest Positioning It is necessary to set lever into its working position see Manipulati on with levers Fig Thighrest positioning Fig Raising the thighrest...

Page 30: ...ned mechanically using raster 1 Mattress holder Calfrest handle 2 Mechanical raster Raise the calfrest as follows Grab the calfrest handle 1 Pull the handle upwards until desired position is reached E...

Page 31: ...osition Ensure there are no body parts between siderails and backrest The bed allows quick mechanical lowering of the backrest using quick counterclockwise rotatition of the right lever Backrest will...

Page 32: ...patient falling out of bed Ensure that folded up siderails are securely anchored Press against siderails from inside in order to check stability Hospital personnel are responsible for ensuring siderai...

Page 33: ...lick when locked in place Ensure that the siderail is locked properly 13Castor Control and Bed Transport WARNING Material damage due to incorrect transport or involuntary movement Prior to assembly di...

Page 34: ...g Maximum load per rail is 10 kg Accessories suitable for accessory rail Urine bag holder Urine bottle holder Cannula holder Radon bottle holder Fig Accessory rail Castors are unbraked Castors are bra...

Page 35: ...or Praktika 1 Accessory rail Lifting pole Infusion stand Protector Extender 16 1Bushings for Accessories Bushings for accessories are placed in the corners of bed head end Following accessories can be...

Page 36: ...acement 2 Backrest 3 Lifting pole in working position top view 4 Lifting pole in working position side view 5 Grab handle Lifting pole innstallation Insert lifting pole 3 4 to one bushings 1 in the he...

Page 37: ...nfusion stand side view 16 3Infusion Stand WARNING Risk of injury or collision when placing other accessories than described in the instructions for use Use infusion stands exclusively for accessories...

Page 38: ...ctor bushing 3 1 Releasing button 2 Tube of Protector 3 Safety pin 4 Bushing of Protector Fig Protector description and installation Install Protector to closed position as follows Insert tube of Prot...

Page 39: ...of injury or collision due to incorrect use Extender can be used only with single collapsible siderails Manufacturer will held no responsibility due using the Extender with different types of siderai...

Page 40: ...e bed condition Check if used cleaning agents and disinfectants are compatible with materials that the product consists of For information see the following table BED COMPONENTS THAT ARE INTENDED TO B...

Page 41: ...ing patient or complete Let the bed dry completely after cleaning or disinfection process After drying the bed place the mattress back on the mattress support platform After drying the bed check funct...

Page 42: ...ufacturer of the detergents according to the material table below Part of bed Praktika 1 Daily C D Changing patient C D Complete C D Stickers Split plastic siderails Stickers Single collapsible sidera...

Page 43: ...echanism Remove the object Adjusting siderail is not possible The siderail lock is dirty Clean the locking mechanism Defective siderail lock Notify service department Faulty brakes The brakes are bloc...

Page 44: ...ts by the original spare parts Check regularly that all accessories are working properly Replace damaged accessories immediately 19 2Spare Parts The serial label is located on the frame of the mattres...

Page 45: ...s of noise and vibration in protected interior spaces of buildings according to stan dard IEC 60601 2 52 Used packaging materials are in accordance with requirements of the Packaging Act Z kon o obale...

Page 46: ...l material and manufacturing re lated failures and errors Failures and errors caused by incorrect use and external effects are not covered Justified complaints will be fixed free of charge during the...

Reviews: