D9U003VA0-0101_04
34
Problem
Symptom Air2Care10
Symptom Air2Care20
Action
Fail to inflate or soft mattress
Low pressure indicator on
NOTE:
This may happen
during normal use while the
mattress is adjusting and no
action is required unless the
indicator remains on for an
extended period.
Indicator error code flashing
twice.
Check air connection to
mattress is ok.
Check that the mattress has
been rolled out flat and air pi
-
pes are not twisted or trapped
in any part of the bedframe.
Open the mattress cover and
check that no air pipes are
damaged or disconnected.
NOTE:
For Air2Care 10 only:
Slightly increase the pressure
setting and see if problem
stops.
Then
1
.
Not alternating
Air cells not alternating in APT
mode
Indicator error code flashing
three times.
Check all air connections
same as for ‘Fail to inflate’
and
1
.
Hard mattress
Air cells very hard in APT or
Static modes
Indictor error code flashing
four times.
Check all air connections
same as for ‘Fail to inflate’.
Reduce pressure setting to
lowest level then
1
.
1
Restart unit by turning power off and then back on. If fault re-occurs turn off SCU and immediately call your local appro
-
ved service provider.
NOTE:
If any of the above problems occur and cannot be solved by the user it is recommended to put the mattress into Transport
mode (see chapter „Transport mode“) to ensure that the patient is supported while waiting for a service engineer to arrive.
For patients who are not bed-ridden, it is possible to connect a seat cushion
operating in a 2-cell cycle to the SCU instead of the mattress. The Air2Care
seat cushion fits on most standard chairs and chairs for specialist geriatric
care. The seat cushion consists of two sets of alternating cells on a foam base
with a front support. The cover consists of vapour-permeable, water-imper
-
meable two-way stretch material. The cushion is only designed for LINET®
SCUs.
Fig. SCU and cushion
9.7 Seating System
9.7.1 Function
The seating system provides alternating pressure therapy for patients that are not bed-ridden.
Have the seating system used only:
■
under supervision of trained and qualified nursing personnel
■
by trained and qualified nursing personnel
Seating system:
■
6 cells
■
cells are inflated and deflated in 12 minutes cycles