Problem

Symptom Air2Care10

Symptom Air2Care20

Action

Fail to inflate or soft mattress 

Low pressure indicator on 

NOTE: 

This may happen 

during normal use while the 

mattress is adjusting and no 

action is required unless the 

indicator remains on for an 

extended period.

Indicator error code flashing  

twice.

Check air connection to 

mattress is ok.

Check that the mattress has 

been rolled out flat and air pi

-

pes are not twisted or trapped 

in any part of the bedframe.

Open the mattress cover and 

check that no air pipes are 

damaged or disconnected. 

NOTE: 

For Air2Care 10 only: 

Slightly increase the pressure 

setting and see if problem 

stops.

Then 

1

.

Not alternating 

Air cells not alternating in APT 

mode 

Indicator error code flashing 

three times.

Check all air connections 

same as for ‘Fail to inflate’ 

and 

1

.

Hard mattress 

Air cells very hard in APT or 

Static modes 

Indictor error code flashing 

four times.

Check all air connections 

same as for ‘Fail to inflate’.

Reduce pressure setting to 

lowest level then 

1

.

1

 

Restart unit by turning power off and then back on.  If fault re-occurs turn off SCU and immediately call your local appro

-

ved service provider.

NOTE: 

If any of the above problems occur and cannot be solved by the user it is recommended to put the mattress into Transport 

mode (see chapter „Transport mode“) to ensure that the patient is supported while waiting for a service engineer to arrive. 

For patients who are not bed-ridden, it is possible to connect a seat cushion 

operating in a 2-cell cycle to the SCU instead of the mattress. The Air2Care 

seat cushion fits on most standard chairs and chairs for specialist geriatric 

care. The seat cushion consists of two sets of alternating cells on a foam base 

with a front support. The cover consists of vapour-permeable, water-imper

-

meable two-way stretch material. The cushion is only designed for LINET® 

SCUs.

Fig. SCU and cushion

9.7 Seating System

9.7.1 Function

The seating system provides alternating pressure therapy for patients that are not bed-ridden.

Have the seating system used only:

■   

under supervision of trained and qualified nursing personnel

■   

by trained and qualified nursing personnel

Seating system:

■   

6 cells

■   

cells are inflated and deflated in 12 minutes cycles

Summary of Contents for Air2Care 10

Page 1: ...Air2Care Model 10 20 Mattress Replacement and Overlay Systems D9U003VA0 0101 Version 04 Publication date 2020 05 Instructions for Use and Technical Description...

Page 2: ...0 Translation LINET 2020 All trademarks and brands are the property of the appropriate owners The manufacturer reserves the right to changes in the contents of the instructions for use that relate to...

Page 3: ...20 6Use and Storage Conditions 21 6 1Storage 22 7Scope of Delivery and Bed Variants 22 7 1Delivery 22 7 2Contents 22 8Putting into Service 23 8 1Securing Strap System 23 8 1 1Installation Overlay Matt...

Page 4: ...arns about the risk of material damage WARNING warns about the risk of physical injury DANGER warns about the risk of fatal injury 1 1 2Structure of Warning Notices SIGNAL WORDS Type and source of dan...

Page 5: ...E WITH CARE THIS WAY UP KEEP DRY PROTECT FROM HUMIDITY PAPER RECYCLING SYMBOL DO NOT USE HAND TRUCK HERE DO NOT STACK DURING STORAGE MEDICAL DEVICE COMPATIBLE WITH MEDICAL DEVICE REGULATION MANUFACTUR...

Page 6: ...NTAINING PHENOL DO NOT WRING REGULARY INSPECT THE INSIDE OF THE COVER FOR CONTAMINATION MACHINE WASH AT 71 C TUMBLE DRY ON LOW HEAT SETTING HANDWASH WITH DETERGENT INITIAL TEMPERATURE OF HOT WATER SHO...

Page 7: ...USE AIR2CARE 20 PROTECTION AGAINST ACCIDENTS DUE TO ELECTRICAL CURRENT TYPE B APPLIED PARTS DOUBLE INSULATION POWER SWITCH I ON O OFF ALTERNATING CURRENT WEEE SYMBOL RECYCLE AS ELECTRONIC WASTE DO NOT...

Page 8: ...Fig Serial label for Air2Care 10 100 127 V 1 Address of Manufacturer 2 1D Bar Code GS1 128 Serial Number 3 PI Product Identifier 4 DI Device Identifier GTIN Global Trade Item Number 5 Manufacturing D...

Page 9: ...l Trade Item Number 5 Manufacturing Date Year Month Day 6 2D Bar Code GS1 DataMatrix DI PI UDI 7 Serial Number 8 Symbols Serial Label for Air2Care 20 SCU 1VAKE20000020 REF SN 20200122672 2020 02 21 01...

Page 10: ...h Label mattress 7 3 2 5 4 6 1VAM000220200 REF SN 20200112325 2020 02 24 01 8592654070567 11 200224 21 20200112325 1 1 Address of Manufacturer 2 1D Bar Code GS1 128 Serial Number 3 PI Product Identifi...

Page 11: ...ure pressure fault 1 8Abbreviations AC Alternating Current CE European Conformity CPR Cardiopulmonary Resuscitation dB Sound Intensity Unit DC Direct Current CUC Configuration number EMC Electromagnet...

Page 12: ...and power supplies WARNING This medical device is not intended for oxygen enriched environment WARNING This medical device is not intended for use with flammable substances WARNING This medical device...

Page 13: ...at installation is performed in accordance with the instructions in the instructions for use Ensure that maintenance is performed only by qualified personnel who have been trained by the manufacturer...

Page 14: ...cm BMI 17 at moderate risk levels in the intensive care Application Environment 1 and 2 as in IEC 60601 2 52 Patients with any stage category of pressure ulcers 3 2Contraindications Mattress system is...

Page 15: ...full length side formers Mattress system cover is water proof and vapour permeable cover made of polyurethene coated nylon which has a two 180 zip It has straps to prevent the mattress ends from movin...

Page 16: ...bottom part non flexible resistant material top part flexible water proof material Mattress replacement system air foam 4 1 3Mattress replacement system air air The mattress replacement system is desi...

Page 17: ...ient s weight distribution or position The SCU is equipped with an audio and or visual alarm system to detect power failures air pipe disconnections or other faults Fig SCU for Air2Care 10 Fig SCU for...

Page 18: ...inflated 7 5 kg 6 kg 2 1 kg 3 5 kg 2 5 kg 2 cells 12 minutes 2 cells 12 minutes Environmental conditions Temperature Humidity Atmospheric pressure 10 C 40 C 30 75 795 1060 hPa Maximal load SWL Mattre...

Page 19: ...ty of this air mattress system and lead to improper operation WARNING Mobile RF communication device including end use devices like antenna cables and external antenna should not be used closer than 3...

Page 20: ...tions equipment IEC 61000 4 3 3 V m 80 MHz 2 7 GHz 80 AM at 1 kHz See Table 1 Fast electrical transients burst IEC 61000 4 4 2 kV for power line repetition frequency 100 kHz Surge IEC 61000 4 5 1 kV L...

Page 21: ...5 500 5 785 5 100 5 800 WLAN 802 11 a n Pulse modulation 217 Hz 9 NOTE There are applied no deviations to requirements of IEC 60601 1 2 ed 4 NOTE There are no known other measures for keeping the basi...

Page 22: ...suitable cover Store in a place suitable for medical devices NOTE Never store more than 4 mattresses on each other This may result by damaging the mattress or system malfunction 7Scope of Delivery and...

Page 23: ...positioned at the Head Thigh and Foot areas of the mattress on both left and right sides Two separate straps are supplied also Each strap has a trident buckle arrangement with a fixed fema le part 1...

Page 24: ...lated Fig Fixing of the mattress 8 2Installation of SCU System Control Unit WARNING Risk of injury when putting SCU into service Make sure that your hands are not trapped between hook and foot board w...

Page 25: ...e sure the air connector is not in transport mode see Transport mode Insert air connector 1 in the socket 2 at an angle of approx 45 Push air connector 1 down until it clicks into place and air connec...

Page 26: ...ing on SCU Disconnect mains cable from mains Fig Switching on SCU 8 4 2Inflation Inflation After switching ON the mattress will start to inflate the indicator is illuminated during inflation Set the p...

Page 27: ...o lack of cleaning Ensure that no moisture gets into the mattress Ensure that no body fluids get into the mattress cover Regulary inspect the inside of the cover for contamination Mattress must be cle...

Page 28: ...TATIC Fully inflated mattress Static mode APT Alternating pressure therapy Dynamic mode APT Alternating pressure therapy Air2Care operates by alternating pressure in a two cell system in a 12 minute i...

Page 29: ...ee System errors 3 Mains switch I On O Off 4 Mode selection button APT Alternating pressure therapy Dynamic mode MAX Maximum inflation mode CLP Constant low pressure mode 5 Mode indicator Indicates th...

Page 30: ...d 9 2 3Pressure Control WARNING Risk of injury due to incorrect pressure setting Consult qualified hospital staff prior to adjusting pressure The recommended pressure levels may not be the optimum for...

Page 31: ...reater comfort then the pressure level can be reduced by turning the pressure control dial to the left then repeating the patient s sacral check Pressure Settings Changes in Patient s Position When a...

Page 32: ...or next to MAX mode stops flashing Switch off SCU via mains switch O Disconnect SCU from mains Mattress is now in Transport mode Option 2 Disconnect air connector from the SCU and twist the end to sea...

Page 33: ...mes Pressure fault Beep Flashes four times Problem Symptom Air2Care10 Symptom Air2Care20 Action Power failure SCU will not turn on Mains switch not illuminated No indicators on front panel illuminated...

Page 34: ...1 1 Restart unit by turning power off and then back on If fault re occurs turn off SCU and immediately call your local appro ved service provider NOTE If any of the above problems occur and cannot be...

Page 35: ...ase for the dynamic cushion allowing it to operate effectively and preventing the patient damaging the cushion when siting in the chair sitting down or standing up Place the Air2Care seat cushion dire...

Page 36: ...ient Ensure that seat cushion is completely inflated Preparation Inflate seat cushion see Installation and Startup Seating of the patient Sit patient on seat cushion For an ideal sitting position Ensu...

Page 37: ...infection control Mattress parts to be cleaned Recommended Cleaning Agents General cleaning Top Cover Seating System Standard hospital detergents Alcohol or Quaternary Ammoni um based disinfectants C...

Page 38: ...before use Even if the cleaner instruc tions say that this is not required This prevents a build up of chemicals on the mattress surface that could reactivate during use and affect biocompatibility C...

Page 39: ...gles 1 on each corner Mattress replacement Air Foam 4 toggles 1 on each corner holding top deck foam is loose Mattress replacement Air Air 4 toggles 1 on each corner holding both air decks Undo next t...

Page 40: ...utside of mattress and outside of SCU for mechanical damage and signs of severe wear and tear Check if mattress and SCU are working properly Check external air filter in side of SCU for dust and dirt...

Page 41: ...ng through an authorized company more details on www linet cz 12 2Disposal The main objective of the obligations arising from the European Directive No 2012 19 EU on Waste Electric and Electronic Equi...

Page 42: ...aterial and manufacturing re lated failures and errors Failures and errors caused by incorrect use and external effects are not covered Justified complaints will be fixed free of charge during the war...

Reviews: