9.3 CPR – Cardipulmonary resuscitation
Fig. CPR Function
9.4 Transport mode
If the SCU becomes disconnected from mains power supply, the system will automatically enter Transport mode and the mattress
will remain inflated but all active pressure adjustment will stop. It is not possible to use any mode in transport mode. The mattress
will stay infalted for up to 12 hours, depending on the patient’s weight and set pressure level when switching into transport mode.
To activate transport mode:
Option 1:
►
For A2C 10 select STATIC mode and move the pressure control dial to its ‚Hi‘ position and wait until the mattress is fully
inflated.
►
For A2C 20 select MAX mode and wait until the indicator (next to MAX mode) stops flashing.
►
Switch off SCU via mains switch (
O
).
►
Disconnect SCU from mains.
Mattress is now in Transport mode.
Option 2:
►
Disconnect air connector from the SCU and twist the end to seal so symbols
and
are heading each other.
Mattress is now in Transport mode.
To deactivate transport mode:
Option 1 and 2:
►
Repeat the process above backwards.
Before starting resuscitation with SCU connected:
►
Press red CPR button.
►
Remove air connector plug from SCU.
►
Mattress starts to deflate and resuscitation procedure can commence.
The air in the mattress will deflate immediately after pressing patient’s chest.
9.5
Power failure
WARNING
Risk of injury due to power failure!
►
Seek clinical advice immediately as alternating pressure therapy is not possible during power failures.
In case of power failure the mattress will remain inflated for up to 12 hours. No active modes are available without mains power.